3-8
250cx Series Maternal/Fetal Monitor
Revision C
2036947-001
Installation: Self-Test Routine
Battery-Backed RAM Status
Whenever you turn off a 250cx Series Monitor, a battery provides power to the
RAM (random access memory) that stores information such as time, date, default
settings, etc.
Low Battery Icon
The icon shown above will appear in the upper right-hand section of the monitor
under the following circumstances.
NOTE:
It is very important to set the time and date prior to initial operation of the
monitor and during daylight-saving time changes. A long-lasting battery
maintains the settings even when the monitor is unplugged from AC
power. Information on setting the time and date is found in the
General
Setup
screen section in Chapter 4 of the
Corometrics 250cx Series Monitor
Operator’s Manual.
Recorder Test
The message TEST: ARE ALL DOTS PRINTED? prints followed by two vertical lines which should
appear continuous. Discontinuous lines may be an indication of damaged printhead elements if gaps
occur in the same place on both lines.
Counting Test
After the recorder test, the display returns to the main screen. The software generates a 120 bpm rate
in the FHR1 area, a 180 bpm rate in the FHR2 area, and both mode titles display Test.
Uterine Activity
The monitor adds 50 mmHg to the previously displayed value in the UA display area; the mode title
displays Test.
Table 7. Monitor Self-Test Routines
Test Routine
Description
Table 8. Battery-Backed RAM Status
Icon Appearance
Reason
Solution
Icon appears and then
disappears after power
cycle.
Data corruption. Your monitor has reverted to
factory settings.
Access setup screens and configure last-used
settings.
Icon appears after most
or all power cycles.
Battery requires service.
Call GE Service to report.
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 6: ...CE CE ii 0086 ...
Page 14: ...viii 250cx Series Maternal Fetal Monitor Revision C 2036947 001 ...
Page 16: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036947 001 2 Equipment Overview ...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 55: ...3 1 250cx Series Maternal Fetal Monitor Revision C 2036947 001 3 Installation ...
Page 56: ...3 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 81: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036947 001 4 Maintenance ...
Page 82: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 141: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036947 001 5 Troubleshooting ...
Page 142: ...5 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 194: ...6 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 237: ...Revision C 250cx Series Maternal Fetal Monitor B 1 2036947 001 B Alarms Summary ...
Page 238: ...B 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Alarms Summary For your notes ...
Page 251: ......