Revision C
250cx Series Maternal/Fetal Monitor
4-51
2036947-001
Maintenance: Power Supply Voltages—Verification
Isolated Power Supply Board Voltages
Verify the following voltages, being sure to use the correct ground reference points.
Isolated FECG/UA Board Voltages
This procedure performs an adjustment on the FECG/UA Board.
Connect the positive lead of a digital voltmeter to TP1 on the FECG/UA Board.
Connect the negative lead to TP2 or TP3 (isolated ground).
Adjust R28 for a reading of +4.00 Vdc ± 0.01 Vdc.
Table 13. Main Board Power Supply Voltages
2023111-001
Signal Name
Voltage Level
Pin 1
+12EL
+12 Vdc ± 0.5 Vdc
Pin 2
+20I
+20 Vdc ± 0.5 Vdc
Pin 3
+15BP
+15 Vdc ± 0.5 Vdc
Pin 4
–15V
–15 Vdc ± 0.5 Vdc
Pin 5
+15V
+15 Vdc ± 0.5 Vdc
Pin 6
+12A
+12 Vdc ± 0.5 Vdc
Pin 7
+5V
+5 Vdc ± 0.5 Vdc
Pin 8
GND
—
Pin 9
No Connection
—
Pin 10
Keying
—
Ground Reference
Test Point
Measurement Limit
TP8
TP7
16.5
±
.15 Vdc
TP8
TP17
15
±
.75 Vdc
TP8
TP16
-15
±
.75 Vdc
TP8
TP18
5
±
.25 Vdc
TP5
TP10
15
±
.75 Vdc
TP5
TP11
-15
±
.75 Vdc
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 6: ...CE CE ii 0086 ...
Page 14: ...viii 250cx Series Maternal Fetal Monitor Revision C 2036947 001 ...
Page 16: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036947 001 2 Equipment Overview ...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 55: ...3 1 250cx Series Maternal Fetal Monitor Revision C 2036947 001 3 Installation ...
Page 56: ...3 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 81: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036947 001 4 Maintenance ...
Page 82: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 141: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036947 001 5 Troubleshooting ...
Page 142: ...5 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 194: ...6 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 237: ...Revision C 250cx Series Maternal Fetal Monitor B 1 2036947 001 B Alarms Summary ...
Page 238: ...B 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Alarms Summary For your notes ...
Page 251: ......