5-36
250cx Series Maternal/Fetal Monitor
Revision C
2036947-001
Troubleshooting: FAQs
How do I check the settings of the Main
CPU Board dip switch SW1 without
opening the unit?
1. Enter the Install Options Screen 1. (
“How do I access the service setup /
2. In the Install Options Screen 1, rotate the Trim Knob to highlight the Test softkey at
the bottom of the display below the menu bar. This opens the Diagnostic Control
screen.
3. At the bottom of the Diagnostic Control screen, the field SW1 appears followed by a
total of 8 ones and zeros. These represent the state of the dip switch settings on the
motherboard. The dip switch positions are left to right, Switch 8 to Switch 1.
A 1 indicates the switch is open (Off), and the associated signal is enabled.
A 0 indicates the switch is closed (On), and the associated signal is disabled.
4. The eight switches are defined as in the following table.
5. On the line immediately below the dip switch settings, the current configuration is
interpreted as text information. For instance, when the dip switch settings indicate
11101100, the configuration line will read “MECG MSpO
2
NIBP CORO “to
indicate that MECG, MSpO
2
and NIBP are enabled, and that the J102 outputs
voltage levels are set to the Corometrics Analog Interface Standard.
6. To exit the Service mode following the instructions at the top of Page 16.
Question
Answer
Switch Positions
Description
Text Interpretation
SW1
Factory Test
TEST or <blank>
SW2
J102 Outputs
CORO or <blank>
SW3
Unused
<blank>
SW4
Clock Speed 16/20 MHz
NIBP or <blank>
SW5
Unused
<blank>
SW6
NBP Option
NBP or <blank>
SW7
MSpO
2
Option
MSpO
2
or
<blank>
SW8
MECG Option
MECG or <blank>
Summary of Contents for Corometrics 250cx Series
Page 2: ......
Page 6: ...CE CE ii 0086 ...
Page 14: ...viii 250cx Series Maternal Fetal Monitor Revision C 2036947 001 ...
Page 16: ...1 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 29: ...Revision C 250cx Series Maternal Fetal Monitor 2 1 2036947 001 2 Equipment Overview ...
Page 30: ...2 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 55: ...3 1 250cx Series Maternal Fetal Monitor Revision C 2036947 001 3 Installation ...
Page 56: ...3 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 81: ...Revision C 250cx Series Maternal Fetal Monitor 4 1 2036947 001 4 Maintenance ...
Page 82: ...4 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 141: ...Revision C 250cx Series Maternal Fetal Monitor 5 1 2036947 001 5 Troubleshooting ...
Page 142: ...5 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 194: ...6 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 For your notes ...
Page 237: ...Revision C 250cx Series Maternal Fetal Monitor B 1 2036947 001 B Alarms Summary ...
Page 238: ...B 2 250cx Series Maternal Fetal Monitor Revision C 2036947 001 Alarms Summary For your notes ...
Page 251: ......