Action
Using the pushable rotary knob to active the
screen.
If the screen can’t be reactivated
then restart the system. If that still does
not help contact the G.A.S. service hotline.
Symtom
Gasflow set-values cannot be achieved.
Possible Cause
Backpressure to low
Action
Set up the back pressure at least to 3 bar.
If that still does not help contact the
G.A.S. service hotline.
Possible Cause
Hardware failure
Action
Contact the G.A.S. service hotline.
Symtom
Temperature set-values cannot be achieved.
Possible Cause
Hardware failure
Action
Contact the G.A.S. service hotline.
Symtom
No display of measurement values during monitoring
Possible Cause
Hardware failure
Action
Contact the G.A.S. service hotline.
Symtom
No Reaction Ion Peak (RIP) will be displayed.
Possible Cause
Device is in negative mode
Action
Switch the drift voltage into positive mode.
If that still does not help contact the
G.A.S. service hotline.
Possible Cause
Hardware failure
Action
Contact the G.A.S. service hotline.
Symtom
No or small Reaction Ion Peak (RIP) will be displayed.
Possible Cause
Insufficient gas quality
Action
Check quality of operating gas (5.0 or
better) and use moisture trap.
Summary of Contents for GC-IMS
Page 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Page 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Page 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Page 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...
Page 212: ...GC IMS User Manual 212 214 Tutorials VOCal Manuals VOCal...