GC-IMS
– User Manual
111/214
8
Connect the Driftgas-/
Carriergas-Adapter to the
Driftgas- and the
Carriergas-Connector.
9
Connect molecular sieve
outlet to
Diftgas-/
Carriergas-Adapter
using a PFA tube.
10
Separately connect
sockets
Sample Out
and
Gas Out
to appropriate
waste gas ventilation
system.
INFORMATION!
To ensure correct measurements it is absolutely necessary to
connect the supplied exhaust tubes
(Gas out and Sample gas out).
The exhaust tubes (Gas out and Sample gas out)
must be led
separately into the exhaust system and must not be connected.
The exhaust system must
not generate any negative pressure
.