GC-IMS
– User Manual
193/214
9.7 Troubleshooting
9.7.1 Error message list
Error message
Drift voltage supply.
Description
Drift voltage error
Action
Fatal Error!
Contact the G.A.S. service
hotline
.
Error message
Can’t save measurement
.
Description
The measurement file could not be saved to the internal
memory.
Action
Export and save all Measurement of the
internal storage and clear the storage.
Repeat the procedure.
If that still does not
help contact the G.A.S. service hotline.
Error message
smb connection failed:
Description
Samba (Service Message Block SMB) connection failed.
Action
Check network cable, network shares and IP-
address. Repeat the procedure.
If that still
does not help contact the G.A.S. service
hotline.
Error message
sftp connection failed:
Description
Secure File Transfer Protocol
(SFTP) connection failed.
Action
Check network cable, network shares and IP-
address. Repeat the procedure.
If that still
does not help contact the G.A.S. service
hotline.
Error message
No USB-Storage found.
Description
Mounting of the USB-Stick failed.
Action
Ensure that a USB-Stick is insert.
Summary of Contents for GC-IMS
Page 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Page 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Page 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Page 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...
Page 212: ...GC IMS User Manual 212 214 Tutorials VOCal Manuals VOCal...