GC-IMS
– User Manual
205/214
Action
Restart the system. If that still does not
help contact the G.A.S. service hotline.
Symtom
Six-Port-valve switching noise sounds strange
Possible Cause
Initialization lost
Action
Reintializise the Six-Port-Valve
Symtom
No sample signals are displayed during a measurement.
Possible Cause
Sample-Pump has failed
Action
Set pump capacity to 100%. The pump device
must be audible loud. Turn the pumping power
back to the working flow. In addition, the
flow can be measured using a flowmeter at the
sample inlet and outlet. If that still does
not help change sample pump.
Symtom
Drift time fluctuates
Possible Cause
System is untight
Action
Check all gas supply connections of operating
gas.
Action
Check all gas supply connections of GC-
column.
Symtom
Retention time fluctuates
Possible Cause
System is untight
Action
Check all gas supply connections of operating
gas.
Action
Check all gas supply connections of GC-
column.