GC-IMS
– User Manual
100/214
INFORMATION!
Only use
stainless steel pressure reducers
,
PFA tubes
with 3 mm
outer diameter and
3 mm Swagelok
connectors.
Screw on the Swagelok caps manually and fix them by screwing a
further half-turn using a 12 mm open-end wrench.
WARNING!
Do not introduce aggressive gases or liquids into the device.
7
Connect nitrogen gas
(Purity 5.0 or better)
source to input of
supplied molecular sieve
using a PFA tube.
INFORMATION!
Recommended pressure of
Operation Gas
(Driftgas / Carriergas) is
3
bar (300 kPa
). 6 bar (600 kPa) must not be exceeded.
8
Connect the
Driftgas-/
Carriergas-Adapter
to
the
Driftgas-
and the
Carriergas-Connector
.
9
Connect molecular sieve
outlet to
Driftgas-/
Carriergas-Adapter
using a PFA tube.
Summary of Contents for GC-IMS
Page 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Page 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Page 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Page 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...
Page 212: ...GC IMS User Manual 212 214 Tutorials VOCal Manuals VOCal...