Error message
No measurements stored.
Description
No measurements stored on the internal storage.
Action
Do a measurement. Repeat the procedure.
If
that still does not help contact the G.A.S.
service hotline.
Error message
Folder cannot be created :
Description
When exporting the device subfolder could not be created.
Action
Check the external storage.
Check network
cable, network shares and IP-address. Check
write protection of the USB-Stick. Ensure
that the USB-Stick is FAT32 formatted.
Reformat the USB-Stick. Replace the USB-Stick
with a new one. Repeat the procedure.
If that
still does not help contact the G.A.S.
service hotline.
Error message
No measurement files transferred
Description
No measurement are transferred to an external storage.
Action
Check the external storage.
Check network
cable, network shares and IP-address. Check
write protection of the USB-Stick. Ensure
that the USB-Stick is FAT32 formatted.
Reformat the USB-Stick. Replace the USB-Stick
with a new one. Repeat the procedure.
If that
still does not help contact the G.A.S.
service hotline.
Error message
No measurements stored.
Description
No measurements were deleted, because no measurements
are available.
Summary of Contents for GC-IMS
Page 1: ...User Manual GC IMS G A S Gesellschaft f r analytische Sensorsysteme mbH...
Page 150: ...GC IMS User Manual 150 214 4 Device without top cover...
Page 164: ...GC IMS User Manual 164 214 17 Tighten the three screws of the frontplate...
Page 188: ...GC IMS User Manual 188 214 9 4 Technical drawing Internal Gasflow...
Page 212: ...GC IMS User Manual 212 214 Tutorials VOCal Manuals VOCal...