Connect µTD device to
power supply.
Connect power supply to
power socket.
INFORMATION!
Only use
stainless steel pressure reducers
,
PFA tubes
with 3 mm
outer diameter and
3 mm Swagelok
connectors.
Screw on the Swagelok caps manually and fix them by screwing a
further half-turn using a 12 mm open-end wrench.
WARNING!
Do not introduce aggressive gases or liquids into the device.
8
Connect nitrogen gas
(Purity 5.0 or better)
source to input of
supplied molecular sieve
using a PFA tube.
INFORMATION!
Recommended pressure of
Operation Gas
(Driftgas / Carriergas) is
3
bar (300 kPa
). 6 bar (600 kPa) must not be exceeded.