Output Pressure Linearity: 0.05%
•
Operation flow rates: 0
– 500 mL/min
Flowcontrol EPC2
Carriergas
•
Type: Differential pressure control
•
Input Pressure: 3.0 bar (300 kPa)
– 6.0 bar
(600 kPa)
•
Output Pressure Stability: 0.01%
•
Output Pressure Linearity: 0.05%
•
Operation flow rates: 0
– 150 mL/min
(depending on the GC-Column dimensions)
Consumables
•
Gas nitrogen 5.0 quality or Synthetic air 5.0
quality
Cleaning mode
•
IMS, column and sampling system are
heated up to > 80°C (~ 100 °C).