GC-IMS
– User Manual
77/214
Snapshot
The
Manage
Button opens the snapshot
window. User-specific target values can
be set. A new snapshot can be recorded
or can be imported
Snapshot is a new function to allow
automatic instrument performance
checking
Contamin
Define the automatic search area for
contamination. The search window can
be defined by setting the Threshold
(signal level in mV which is recognised as
a contamination) and the Start of the
search in the drift-time spectrum given as
RIP relative proportional factor.
•
Current:
Shows the currently
detected contamination Shows the
currently detected contamination.
•
Threshold:
The set value from
which an error is triggered.
Search from:
The RIP-relative starting
position of the search.
Current Loops
(optional)
Activating the
Settings Button
opens a
current loop setup dialog (only visible if
available)
Live Spectra:
Prepared for further use! Currently not in
function!
Live Sensor:
The settings for Live Data communication
with G.A.S. Sequence Designer Software
can be set. For detailed Information refer
to the G.A.S. Sequence Designer
Software Manual.