GC-IMS
– User Manual
165/214
8.13 Workflow: Reinitialize the 6-Port-Valve
INFORMATION!
Only authorized, trained and technically instructed people are
allowed to do these work steps.
DANGER!
The device must be operated in the open state.
Be very careful!
1
Remove the top cover.
For detailed information
see
Workflow: Remove
Housing Cover
2a
Select
Spectra
›
Sample
.
Switch valve
several
times
between position
Fill Loop
and
Inject
.