Revision 1
84
6/27/2017
w) Remove the backpurge by selecting the
Sample
check box.
x) Click the On Line / Off Line switch to indicate that the system is
On Line
. The system will
automatically restart itself. The analyzer may show a fault and be in Manual (back purge)
mode when put On Line. This is due to a low cell temperature. Once the oven
temperature achieves its set point and the preset time delay expires, the system will
automatically restart.
7.5 Measurement Cell Removal
The measurement cell will require removal only when there is a suspected sampling problem
or if it becomes necessary to change the cell windows and/or O-Rings. A sampling problem
could be due to system plugging or collection of nongaseous material in the measurement
cell.
The measurement cell is to be removed by trained and authorized
personnel only. The oven compartment and all of its components are HOT
(150
o
C). Wear appropriate personal protective equipment (gloves, eye wear,
clothing etc) if working with hot surfaces or wait until surface have cooled
to safe temperatures before performing maintenance tasks.
The control cabinet door may NOT be opened while the analyzer is
energized UNLESS the area is known to be non-hazardous. Observe all the
warning labels on the analyzer enclosures.
Procedure
a) Click the On Line / Off Line switch to indicate that the system is
Off Line
.
b) Place the analyzer in back purge mode. This can be done by navigating to the
Back
Purge
field on the Analysis 1 page and pressing
Enter
to place an “x” in the box
The control cabinet door may NOT be opened while the analyzer is
energized UNLESS the area is known to be non-hazardous. Observe all the
warning labels on the analyzer enclosures.
c) Access the purged control cabinet following proper user company and/or regulatory
agency procedure.
d) Turn the UV Source Lamp off by placing the UV Lamp switch to the OFF
p
osition.
e) Turn the oven heaters off by locating the AC Power fuse terminal on terminal strip
ACTS:25. Lift the fuse terminal tab opening the terminal block and thus disconnecting the
AC Power fuse. Refer to the Control Cabinet Components drawing.
f) Remove the oven cover.
Summary of Contents for 943-TGS
Page 50: ...Revision 1 50 6 27 2017 Figure 4 14 Network Sub Panel ...
Page 94: ...Revision 1 94 6 27 2017 Figure 9 1 Mounting Detail ...
Page 95: ...Revision 1 95 6 27 2017 Figure 9 2 Oven Cabinet Detail ...
Page 96: ...Revision 1 96 6 27 2017 Figure 9 3 Control Cabinet Detail ...
Page 97: ...Revision 1 97 6 27 2017 Figure 9 4 Power Air Steam Requirements and Input Output Detail ...
Page 98: ...Revision 1 98 6 27 2017 Figure 9 5 Control Cabinet to Oven Cabinet Interconnect Detail ...
Page 99: ...Revision 1 99 6 27 2017 Figure 9 6 AC Wiring Detail ...
Page 100: ...Revision 1 100 6 27 2017 Figure 9 7 DC and Signal Wiring Detail ...
Page 101: ...Revision 1 101 6 27 2017 This page intentionally left blank ...