6.4 Optimizing the Spectrometer Energy Level
Optimization of the Spectrometer Energy Level is periodically required to maintain a sufficient
signal to noise ratio on the spectrometer output signal as the spectral output signal from the
spectrometer will deteriorate over time. The normal causes of this deterioration are:
•
Degrading Anti-Solarant Solution
•
Aging UV Source Lamp (output reduces over time)
•
Dirty cell windows
•
Dirty optics
•
Optical fibre problem
For convenience, the energy level of the spectrometer output signal is gauged by observing
the peak level as displayed on the Spectrum panel of analyzer or the Analysis/Spectrum
page on the Web GUI under zero gas conditions. When optimizing the spectrometer signal,
the normal target value for the peak value is 32000 counts (+/- 500).
The spectrometer energy level can be optimized manually (for special cases), or
automatically as a function of the analyzer. The automatic function will always target 32000
counts.
Before proceeding with either procedure below, ensure that the analyzer is
‘Off Line’ and is being purged with zero gas.
6.4.1 Automated Optimization
a) Place the analyzer in back purge mode. This can be done by accessing the
Back Purge
field on the
Analysis 1
panel and pressing
Enter
to place an
X
in the box.
b) Access the
Spectrum
panel, navigate to the
Optimize Integ Time
field and press
Ente
r.
c) Return to the
Analysis 1
Panel.
d) Request a Manual Zero by navigating to the “Manual Zero” field and pressing Enter.
e) Remove the Zero mode by accessing the
Sample
field and then pressing Enter to put an
x in the box, thus allowing the analyzer to return to normal sampling.
6.4.2 Manual Optimization
a) Place the analyzer in Zero mode by accessing the
Analysis 1
panel, place an X in the
Back Purge
field and pressing Enter.
b) Access the
Spectrum
panel.
c) If the peak level is between 25,000 and 32,000 then no adjustment is required at this
time.
d) If adjustment is necessary, optimize the peak signal height to 32,000 counts. Navigate to
the “
Set Integ Time
” field and press
Enter
to display the
Integration Period
dialog box.
Summary of Contents for 943-TGS
Page 50: ...Revision 1 50 6 27 2017 Figure 4 14 Network Sub Panel ...
Page 94: ...Revision 1 94 6 27 2017 Figure 9 1 Mounting Detail ...
Page 95: ...Revision 1 95 6 27 2017 Figure 9 2 Oven Cabinet Detail ...
Page 96: ...Revision 1 96 6 27 2017 Figure 9 3 Control Cabinet Detail ...
Page 97: ...Revision 1 97 6 27 2017 Figure 9 4 Power Air Steam Requirements and Input Output Detail ...
Page 98: ...Revision 1 98 6 27 2017 Figure 9 5 Control Cabinet to Oven Cabinet Interconnect Detail ...
Page 99: ...Revision 1 99 6 27 2017 Figure 9 6 AC Wiring Detail ...
Page 100: ...Revision 1 100 6 27 2017 Figure 9 7 DC and Signal Wiring Detail ...
Page 101: ...Revision 1 101 6 27 2017 This page intentionally left blank ...