Table 6-1: Normal Values of Key Operating Parameters
Inspection Description
Should be:
1
On/Off Line Indicator (Analysis
1 Panel)
•
On Line (green)
2
Run/Stop Indicator
•
Run (green)
3
Relay Indicators (Analysis 1
Panel)
•
Status “Normal” (green)
•
Service “Normal” (green)
•
Mode “Run” (green)
•
Control “Auto” (green)
4
Fault Indicators (Indicators
Panel)
•
Cell Temperature (Green)
•
Probe Temperature (Green)
•
Measure Range (Green)
•
Reference Range (Green)
•
Dark Range (Green)
•
I/O Board (green)
5
Cell Temperature
(Typical Cell Temperature =
150
o
C/302
o
F) (Analysis 2
Panel)
•
Digital Indicator (within set point)
6
Probe Temperature
(Typical Cell Temperature =
150
o
C/302
o
F) (Analysis 2
Panel)
•
Digital Indicator (within set point). This
parameter is used only when a probe RTD is
installed and the Control with RTD function is
selected
7
Cabinet Temperature (Analysis
2 Panel)
•
Digital Indicator
8
Air Demand Output
•
Digital - Indicating -5% to + 5%
•
Graphic Showing trend, last value on right side
9
H
2
S and SO
2
•
Digital – Separate indications in mole %
•
Graphic Dual trace trend, last value on right
side
Summary of Contents for 943-TGS
Page 50: ...Revision 1 50 6 27 2017 Figure 4 14 Network Sub Panel ...
Page 94: ...Revision 1 94 6 27 2017 Figure 9 1 Mounting Detail ...
Page 95: ...Revision 1 95 6 27 2017 Figure 9 2 Oven Cabinet Detail ...
Page 96: ...Revision 1 96 6 27 2017 Figure 9 3 Control Cabinet Detail ...
Page 97: ...Revision 1 97 6 27 2017 Figure 9 4 Power Air Steam Requirements and Input Output Detail ...
Page 98: ...Revision 1 98 6 27 2017 Figure 9 5 Control Cabinet to Oven Cabinet Interconnect Detail ...
Page 99: ...Revision 1 99 6 27 2017 Figure 9 6 AC Wiring Detail ...
Page 100: ...Revision 1 100 6 27 2017 Figure 9 7 DC and Signal Wiring Detail ...
Page 101: ...Revision 1 101 6 27 2017 This page intentionally left blank ...