Table 7-1: Analyzer Fault Table
Cell Temperature
•
Oven heater fuse is blown (ACTS:25)
•
Cell assembly is not securely clamped to the heaters
•
Oven heater(s) is inoperative
•
Set-point for low temperature alarm is incorrect
High High Sulfur
•
The sulfur vapor concentration in the sample has risen
to unacceptable levels. This usually indicates that the
probe’s cold finger is too warm.
Measure Range
Reference Range
•
UV lamp is nearing the end of its useful life (see
Section 7.3)
•
UV lamp is not operating (see Section 7.3)
•
Anti-Solarant needs changing or has been changed
using expired solution
•
Cell windows are dirty
•
Optical Fiber defective
•
UV lamp power supply is blown (ACTS:28)
Dark Range
•
Spectrometer problem, consult factory
.
I/O Board
•
The I/O board is not communicating with the controller
board.
•
Check ArcNet wiring.
Cabinet Temperature
•
RTD problem
•
Computer temperature control is malfunctioning
•
Air supply to Vortec cooler is turned off.
High Absorbance
•
The H
2
S and or SO
2
concentration is out of range.
•
The analyzer is not zeroed.
High Integration Period
•
UV lamp is nearing the end of its useful life (see
Section 7.3)
•
UV lamp is not operating (see Section 7.3)
•
Anti-Solarant needs changing or has been changed
using expired solution
•
Cell windows are dirty
•
Optical Fiber defective
•
UV lamp power supply is blown (ACTS:28)
Low Reference Peak
•
UV lamp is nearing the end of its useful life (see
Section 7.3)
•
UV lamp is not operating (see Section 7.3)
•
Anti-Solarant needs changing or has been changed
using expired solution
•
Cell windows are dirty
•
Optical Fiber defective
•
UV lamp power supply is blown (ACTS:28)
High Sulfur Vapor
•
The sulfur vapor concentration in the cell is
approaching the maximum value. The analyzer
readings are still valid.
Summary of Contents for 943-TGS
Page 50: ...Revision 1 50 6 27 2017 Figure 4 14 Network Sub Panel ...
Page 94: ...Revision 1 94 6 27 2017 Figure 9 1 Mounting Detail ...
Page 95: ...Revision 1 95 6 27 2017 Figure 9 2 Oven Cabinet Detail ...
Page 96: ...Revision 1 96 6 27 2017 Figure 9 3 Control Cabinet Detail ...
Page 97: ...Revision 1 97 6 27 2017 Figure 9 4 Power Air Steam Requirements and Input Output Detail ...
Page 98: ...Revision 1 98 6 27 2017 Figure 9 5 Control Cabinet to Oven Cabinet Interconnect Detail ...
Page 99: ...Revision 1 99 6 27 2017 Figure 9 6 AC Wiring Detail ...
Page 100: ...Revision 1 100 6 27 2017 Figure 9 7 DC and Signal Wiring Detail ...
Page 101: ...Revision 1 101 6 27 2017 This page intentionally left blank ...