Revision 1
108
6/27/2017
12.3.1 Analog Output Fields
Each
Analog Output
box allows the user to configure, calibrate and test an analog output.
12.3.1.1 Calibrating and Testing the Analog Outputs
To test or calibrate any I/O point, first click on the
Direct Peripheral Board
Control
checkbox.
To configure/calibrate/test an Analog Output:
a) Place a check mark in the
Calibrate Analog Output x
check box (the approach is the
same for all four boxes). The fields within the box will become active.
b) Enter the desired Zero and Span values for the output in the
Scaled Value/Zero
and
Span
fields. For example, if the output is to represent 0 - 2 % H2S, enter 0 in the
Zero
field and 2 in the
Span
field.
c) Connect a milli-ammeter to
Analog Output x
. Note that the analog outputs require
external loop power.
d) Press the
Output Zero
button. The analog output will attempt to output 4 mA, however it
may not read exactly 4.
e) Enter the reading from the meter in the
Meter Reading
field and press the
Auto Cal
button. This will adjust the
Zero
output to exactly 4 mA.
f) Press the
Output Span
button. The analog output will attempt to output 20 mA, however
it may not read exactly 20.
g) Enter the reading from the meter in the Meter Reading field and press the
Auto Cal
button. This will adjust the
Span
output to exactly 20 mA. The analog output is now
calibrated.
h) Test the output by entering the desired mA output in the
Test(mA)
field and then pressing
the
Output Test
button. This will force the analog output to generate
the entered value.
In this way the analog output loop can be tested at several readings such as 0, 25, 50, 75
and 100 % of scale (4, 8, 12, 16 and 20 mA).
i) Once the Zero value and Span values have been tested, press the
Capture
button to
implement the changes.
j) Press
Permanent Save
.
Summary of Contents for 943-TGS
Page 50: ...Revision 1 50 6 27 2017 Figure 4 14 Network Sub Panel ...
Page 94: ...Revision 1 94 6 27 2017 Figure 9 1 Mounting Detail ...
Page 95: ...Revision 1 95 6 27 2017 Figure 9 2 Oven Cabinet Detail ...
Page 96: ...Revision 1 96 6 27 2017 Figure 9 3 Control Cabinet Detail ...
Page 97: ...Revision 1 97 6 27 2017 Figure 9 4 Power Air Steam Requirements and Input Output Detail ...
Page 98: ...Revision 1 98 6 27 2017 Figure 9 5 Control Cabinet to Oven Cabinet Interconnect Detail ...
Page 99: ...Revision 1 99 6 27 2017 Figure 9 6 AC Wiring Detail ...
Page 100: ...Revision 1 100 6 27 2017 Figure 9 7 DC and Signal Wiring Detail ...
Page 101: ...Revision 1 101 6 27 2017 This page intentionally left blank ...