52
Acoustic output
Table 14: Transducer model:
rC60xi
Operating mode:
PW Doppler
Index label
M.I.
TIS
TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
<
1
A
aprt
>1
Global maximum index value
1.2
—
—
2.0
4.0
(b)
Associated acoustic
param
e
te
r
p
r.3
(MPa)
1.73
W
0
(mW)
—
—
291.8
#
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
187.5
z
1
(cm)
4.0
Z
bp
(cm)
4.0
Z
sp
(cm)
3.60
z@PII
.3max
(cm)
4.5
d
eq
(Z
sp
)
(cm)
0.94
F
c
(MHz)
2.20
—
—
2.23
2.23
#
Dim of A
aprt
X (cm)
—
—
4.77
3.28
#
Y (cm)
—
—
1.20
1.20
#
O
ther information
PD
(μsec)
1.153
PRF
(Hz)
1302
p
r
@PII
max
(MPa)
2.43
d
eq
@Pll
max
(cm)
0.54
Focal Length
FL
x
(cm)
—
—
17.97
#
FL
y
(cm)
—
—
6.50
#
I
PA.3
@MI
max
(W/cm
2
)
267
Operating
contr
o
l
co
nditio
ns
Control 1: Exam type
Abd
Abd
Abd
Control 2: Sample volume size
3 mm
7 mm
7 mm
Control 3: Sample volume position
Zone 3
Zone 6
Zone 5
Control 4: PRF
1302
2604
2604
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#
No data are reported for this operating condition since the global maximum index value is not reported for the
reason listed. (Reference Global Maximum Index Value
line.)
— Data are not applicable for this transducer/mode.