![FujiFilm SonoSite SII User Manual Supplement Download Page 48](http://html1.mh-extra.com/html/fujifilm/sonosite-sii/sonosite-sii_user-manual-supplement_2343393048.webp)
46
Acoustic output
Output display
Table 10: TI or MI
≥
1.0
Transducer
Index
2D/M Mode
CPD/Color
PW Doppler
CW Doppler
C8x
MI
Yes
Yes
Yes
—
TIC, TIB, or
TIS
No
No
Yes
—
C11x
MI
No
No
No
—
TIC,TIB, or TIS
No
No
Yes
—
C35x
MI
Yes
No
No
—
TIC, TIB, or
TIS
No
No
Yes
—
rC60xi standard/
armored
MI
Yes
Yes
Yes
—
TIC,TIB, or TIS
Yes
Yes
Yes
—
HFL38xi standard/
armored
MI
Yes
Yes
Yes
—
TIC, TIB, or
TIS
No
No
Yes
—
HFL50x
MI
Yes
Yes
Yes
—
TIC, TIB, or
TIS
No
No
Yes
—
HSL25x
MI
Yes
Yes
No
—
TIC, TIB, or
TIS
No
No
Yes
—
ICTx
MI
No
No
No
—
TIC, TIB, or
TIS
No
No
Yes
—
Even when MI is less than 1.0, the system provides a continuous real-time display of MI in all imaging modes, in
increments of 0.1.
The system meets the output display standard for TI and provides a continuous real-time display of TI in all imaging
modes, in increments of 0.1.
The TI consists of three user-selectable indices, and only one of these is displayed at any one time. In order to
display TI properly and meet the ALARA principle, the user selects an appropriate TI based on the specific exam
being performed. FUJIFILM SonoSite provides a copy of
AIUM Medical Ultrasound Safety
, which contains guidance
on determining which TI is appropriate.