64
Acoustic output
Table 26: Transducer
m
odel:
rP19x
Operating mode:
PW Doppler
Index label
M.I.
TIS
TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
≤1
A
aprt
>1
Global maximum index value
1.3
—
—
1.8
4.0
3.9
As
so
ciat
ed
aco
u
st
ic
parameter
p
r.3
(MPa)
1.94
W
0
(mW)
—
—
240.2
251.1
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
173.7
z
1
(cm)
2.5
Z
bp
(cm)
2.5
Z
sp
(cm)
3.35
z@PII
.3max
(cm)
3.0
d
eq
(Z
sp
)
(cm)
0.80
F
c
(MHz)
2.14
—
—
2.23
2.23
2.10
Dim of A
aprt
X (cm)
—
—
1.86
1.80
1.80
Y (cm)
—
—
1.15
1.15
1.15
O
ther information
PD
(μsec)
1.334
PRF
(Hz)
1562
p
r
@PII
max
(MPa)
2.42
d
eq
@Pll
max
(cm)
0.62
Focal Length
FL
x
(cm)
—
—
29.82
18.46
FL
y
(cm)
—
—
9.00
9.00
I
PA.3
@MI
max
(W/cm
2
) 180
Operating
contr
o
l
co
nditio
ns
Control 1: Exam type
Crd
Crd
Crd
Crd
Control 2: Sample volume size
1 mm
12 mm
12 mm
1 mm
Control 3: Sample volume position Zone 1
Zone 7
Zone 5
Zone 5
Control 4: PRF
1562 Hz
1562 Hz 1562 Hz
3906 Hz
Control 5: TDI
Off
Off
Off
Off
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#
No data are reported for this operating condition since the global maximum index value is not reported for the
reason listed. (Reference Global Maximum Index Value
line.)
— Data are not applicable for this transducer/mode.