Acoustic output
59
English
De
ut
sch
Español
Fr
a
n
ç
a
is
Ita
lia
no
P
o
rtuguês
Nederlands
Table 21: Transducer model:
L25x
Operating mode:
PW Doppler
Index label
M.I.
TIS
TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
≤1
A
aprt
>1
Global maximum index value
(a)
—
(a)
—
1.7
(b)
As
so
ciat
ed
acou
st
ic
parameter
p
r.3
(MPa)
#
W
0
(mW)
—
#
32.1
#
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
—
z
1
(cm)
—
Z
bp
(cm)
—
Z
sp
(cm)
0.75
d
eq
(Z
sp
)
(cm)
0.30
F
c
(MHz)
#
—
#
—
6.00
#
Dim of A
aprt
X (cm)
—
#
—
0.76
#
Y (cm)
—
#
—
0.30
#
Other Information
PD
(μsec)
#
PRF
(Hz)
#
p
r
@PII
max
(MPa)
#
d
eq
@Pll
max
(cm)
0.21
Focal Length
FL
x
(cm)
—
#
—
#
FL
y
(cm)
—
#
—
#
I
PA.3
@MI
max
(W/cm
2
) #
Operating
contr
o
l
co
nditio
ns
Control 1: Exam type
—
—
—
—
Vas/Ven/Nrv —
Control 2: Sample volume size
—
—
—
—
8 mm
—
Control 3: Sample volume position —
—
—
—
Zone 7
—
Control 4: PRF
—
—
—
—
1953
—
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#
No data are reported for this operating condition since the global maximum index value is not reported for the
reason listed. (Reference Global Maximum Index Value
line.)
— Data are not applicable for this transducer/mode.