34
Measurement references
g
Using the touchpad or the touchscreen, position the caliper at the end of the desired waveform, and
then tap
Set
.
h
Tap
Save Calc
to save the calculation.
To measure a Tissue Doppler Imaging (TDI) waveform
1
Ensure that TDI is on.
2
On a frozen Doppler spectral trace, tap
Calcs
.
3
Tap
TDI
on the calculations menu, and then do the following for each measurement you want to take:
a
On the calculations menu, select the measurement name.
b
Position the calipers.
c
Tap
Save Calc
to save the calculation.
Measurement references
Measurement accuracy
Note
If you invert the frozen image or move the baseline, results are cleared.
Table 7: PW Doppler mode measurement and calculation accuracy and range
Doppler mode
measurement
accuracy and range
System tolerance
Accuracy by
Test method
a
Range
Velocity cursor
< +/- 2% plus 1% of full
scale
b
Acquisition
Phantom
0.01 cm/sec-
550 cm/sec
Frequency cursor
< +/- 2% plus 1% of full
scale
b
Acquisition
Phantom
0.01kHz-20.8
kHz
Time
< +/- 2% plus 1% of full
scale
c
Acquisition
Phantom
0.01-10 sec
a
FUJIFILM SonoSite special test equipment was used.
b
Full scale for frequency or velocity implies the total frequency or velocity magnitude, displayed on the scrolling
graphic image.
c
Full scale for time implies the total time displayed on the scrolling graphic image.