32
Measurements and calculations
b
Use the automatic trace tool to trace the waveform. See
c
Tap
Save Calc
to save the calculation.
To calculate Stroke Volume (SV) or Stroke Index (SI)
The SV and SI calculations require a measurement in 2D and a measurement in Doppler. SI also requires
Body Surface Area (BSA). After the measurements are saved, the result appears in the patient report.
1
(SI Only) Fill in the
Height
and
Weight
fields on the patient form. The BSA is calculated automatically.
2
Measure from LVOT (2D):
a
On a frozen 2D image, tap
Calcs
.
b
On the calculations menu, tap
Ao/LA
then
LVOT D
.
c
Position the calipers.
d
Tap
Save Calc
to save the calculation.
3
Measure from
LVOT
(Doppler). Refer to
“To calculate Velocity Time Integral (VTI)”
on page 29. On
the calculations menu, tap
AV
and then
LVOT VTI
.
To calculate Cardiac Output (CO) or Cardiac Index (CI)
The CO and CI calculations require Stroke Volume (SV) and Heart Rate (HR) calculations. CI also requires
Body Surface Area (BSA). After the measurements are saved, the result appears in the patient report.
1
(CI Only) Fill in the
Height
and
Weight
fields on the patient form. The BSA is calculated automatically.
2
Calculate SV as described in
“To calculate Stroke Volume (SV) or Stroke Index (SI)”
3
Calculate HR as described in