![FujiFilm SonoSite SII User Manual Supplement Download Page 68](http://html1.mh-extra.com/html/fujifilm/sonosite-sii/sonosite-sii_user-manual-supplement_2343393068.webp)
66
Acoustic output
Table 28: Transducer model:
rP19x
Operating mode:
CW Doppler
Index label
M.I.
TIS
TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
≤1
A
aprt
>1
Global maximum index value
(a)
—
1.2
—
4.0
4.0
As
so
ciat
ed
aco
u
st
ic
parameter
p
r.3
(MPa)
#
W
0
(mW)
—
125.4
125.4
125.4
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
—
z
1
(cm)
—
Z
bp
(cm)
—
Z
sp
(cm)
0.90
z@PII
.3max
(cm)
#
d
eq
(Z
sp
)
(cm)
0.64
F
c
(MHz)
#
—
2.00
—
2.00
2.00
Dim of A
aprt
X (cm)
—
0.42
—
0.42
0.42
Y (cm)
—
1.15
—
1.15
1.15
O
ther information
PD
(μsec)
#
PRF
(Hz)
#
p
r
@PII
max
(MPa)
#
d
eq
@Pll
max
(cm)
0.61
Focal Length
FL
x
(cm)
—
1.55
—
1.55
FL
y
(cm)
—
9.00
—
9.00
I
PA.3
@MI
max
(W/cm
2
) #
Operating
contr
o
l
co
nditio
ns Control 1: Exam type
Crd
Crd
Crd
Control 2: Sample volume position
Zone 0
Zone 0
Zone 0
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#
No data are reported for this operating condition since the global maximum index value is not reported for the
reason listed. (Reference Global Maximum Index Value
line.)
— Data are not applicable for this transducer/mode.