54
Acoustic output
Table 16: Transducer Model:
HFL38xi Ophthalmic Use
Operating Mode:
PW Doppler
Index Label
M.I.
TIS
TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
<
1
A
aprt
>1
Global Maximum Index Value
0.18
—
0.09
—
0.17
(b)
As
so
ciat
ed
A
c
o
u
st
ic
Pa
ra
m
e
te
r
p
r.3
(MPa)
0.41
W
0
(mW)
—
3.56
3.56
#
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
—
z
1
(cm)
—
Z
bp
(cm)
—
Z
sp
(cm)
1.64
z@PII
.3max
(cm)
0.9
d
eq
(Z
sp
)
(cm)
0.31
F
c
(MHz)
5.34
—
5.33
—
5.33
#
Dim of A
aprt
X (cm)
—
1.08
—
1.08
#
Y (cm)
—
0.40
—
0.40
#
Other Information
PD
(μsec)
1.28
PRF
(Hz)
1302
p
r
@PII
max
(MPa)
0.48
d
eq
@Pll
max
(cm)
0.19
Focal Length
FL
x
(cm)
—
3.72
—
#
FL
y
(cm)
—
2.44
—
#
I
PA.3
@MI
max
(W/cm
2
)
6.6
Operating
Contr
o
l
Co
ndi
tio
ns
Control 1: Exam Type
Oph
Oph
Oph
Control 2: Sample Volume Size
1 mm
10 mm
10 mm
Control 3: Sample Volume Position
Zone 1
Zone 7
Zone 7
Control 4: PRF
1302
10417
10417
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
#
No data are reported for this operating condition since the global maximum index value is not reported for the
reason listed. (Reference Global Maximum Index Value
line.)
— Data are not applicable for this transducer/mode.