6.10 Checking alarm output
6.10.1 Low Light Transmission (1 a contact output)
Remove the receiver box or the transmitter box to generate “Low light Transmission” alarm.
At this time, do not look into the transmitter box. Otherwise, it may cause serious damage to your retinae or
cornea. It takes about a minute until “Low Light Transmission” alarm is generated.
If this is done just after the power off, it takes about 6 minutes.
After you confirm the alarm output, attach the receiver box or the transmitter box to the angle adjustment
unit where they were.
6.10.2 Analyzer faulty (1 a contact output)
Turn off the power, and remove the connector of the cable between the receiver unit and the control unit. If
the power is turned on in the state above, “Connection Error” alarm is generated, and “Analyzer faulty”
alarm is then generated.
After you confirm the alarm output, turn off the power, connect the cable, and turn on the power.
6.10.3 On Hold / Under Calibration (1 a contact output)
Hold the output in reference to “6.4 Output hold”. Check the operation output on hold/under calibration. Af-
ter checking it, release hold.
6.10.4 Beyond the upper/lower limits (1 a contact output)
Set the alarm so that the current measured value is beyond the upper or lower limit to generate the alarm, in
reference to “6.3 Alarm setting”.
After checking the operation, set the alarm to OFF or change the setting to the appropriate one.
6.10.5 Power OFF (Closed contact output during power OFF)
Turn OFF the power, and then check the operation output of Power OFF.