![Fuji Electric ZSS-7 Instruction Manual Download Page 80](http://html1.mh-extra.com/html/fuji-electric/zss-7/zss-7_instruction-manual_2341920080.webp)
INZ-TN7ZSS-E
- 73 -
Remote average value reset (DI1 terminal)
Apply rectangular waveform voltage (pulse width: 2.0 sec or more) to the input terminal of remote average
value reset (DI1 terminal, option) to reset the average value.
Reset is kept during short-circuit.
When switching the action from short-circuiting to opening, average value action starts.
6.5.4 Key lock
Key lock disables the settings other than key lock OFF.
When the key lock is turned ON,
Any screens except “Parameter Setting” can-
not be displayed, even if the item is selected
and the
key is pressed on the “Menu”
screen.
Any items except “Key Lock” cannot be vali-
dated, even if the item is selected and the
key is pressed on the “Parameter Setting” screen.
2006-04-01
12:00:00
Menu
Zero Calibration
Alarm Log
Display/Output
Pasword
Parameter Setting
Input Signal
Span Calibration
Alarm Setting
AO Select
Output Hold
Analyzer Information
ENT
2006-04-01
12:00:00
mm
01000
HC
Min.
01
%
12
OFF
OFF
09
07/06/02 16:00:19
Min.
Average Period
cv O
2
Reset Avg.Gas
Key Lock
Disp.OFF
Date/Time
Parameter Setting
Path lengths
2.0s or more
Remote hold signal
OFF
12 to 24 VDC/
5 to 20 mA
Starts average value reset