![Fuji Electric ZSS-7 Instruction Manual Download Page 108](http://html1.mh-extra.com/html/fuji-electric/zss-7/zss-7_instruction-manual_2341920108.webp)
INZ-TN7ZSS-E
- 101 -
1-5
Installation environment
(1)
Ambient temperature
: –20 to 55°C (Receiver unit/Transmitter unit), –5 to 45°C (control
unit)
(2)
Ambient humidity
: 90% R.H. or less
(3)
Measurable optical path
length (stack diameter)
: 0.5 to 10m (CO+O
2
: 0.5 to 5 m)
(4)
Standard flange
: JIS10K 50A flange (JIS B 2212)
(5)
Air purge
: Instrument air (compressor must be installed when power supply
cannot be provided.) (Oil or mist should not be contained), N
2
for
O
2
analyzer excluding air purge version
Pressure 0.5 to 0.7MPa or more 1.0 MPa or less
(6)
Air purge flow rate
: 20L/min or more (depending on the measured gas and velocity)
(7)
Measured gas condition
: Temperature: as specified in “1-1 (4) Measurable component and
range” of APPENDIX 1.
Pressure: ±10kPa
Moisture: 50vol% or less (or should not be saturated water vapor)
Velocity: 25m/s or less (However, consultation is necessary for the
environment where dust (1g/Nm
3
or more) or water (25vol% or
more) exists.) (Prevention of dust deposition or dew condensation
due to increase in air purge flow rate is required)
(8)
Dust
: Standard: 5 g/Nm
3
or less
High-speed/dust-proof/AGC: O
2
analyzer: 10 g/m
3
or less
:
Excluding
O
2
analyzer: 15 g/m
3
(N) or less
(9)
Vibration
: 0.5G or less (0.2G or less when the frequency range is 20 to 40Hz)
(When the optical path length is 1m.)
(10) Mounting angle
: Horizontally ±5 degree or less (No dew condensation should accu-
mulate on the window.)
(11) Light axis fluctuation
range
: Within 0.3 degree (When the path length is 0.5m)
1-6
Performance
(1) Repeatability
: ±1.0% FS (depending on measurable component and range)
*CO+O
2
: ±2.0%FS
(2) Linearity
: ±1.0% FS (depending on measurable component and range)
*CO+O2: ±3.0%FS
(3) Zero drift
: ±2.0%FS / 6 months (depending on measurable component and
range)
*CO+O2: ±4.0%FS
(4) Interference from other
gas components
: ±2.0%FS
(5) Minimum detectable limit : 1% of the minimum range
(6) Response time (90% FS
response)
: 5 seconds or less
(7) Warm up time
: 90 minutes or less
Summary of Contents for ZSS-7