INZ-TN7ZSS-E
- 5 -
Name
Description
(1) Receiver box
A photodiode to receive light and a board are built-in.
(2) Transmitter box
Laser element, peltier to control the temperature of laser and a board are
built-in.
(3) Adjustment probe for received
light quantity
Terminal which checks the level of received light quantity with the voltage
value when light axis adjustment is performed. It is provided in both the
receiver unit and transmitter unit. Connecting the terminal is the BNC
connector.
(4) Angle adjustment unit
A unit which adjusts light axis to receive laser light transmitted from
transmitter unit via PD of receiver unit. Up to 5° can be angle-tuned.
(5) Bolt for angle fine adjustment
When the stack becomes longer, received light quantity changes greatly by
subtle angle-turning. In such cases, light axis adjustment is performed by
using the fine adjustment screw. It may not be necessary when the stack
(optical path lengths) is short. Use it according to the length of the stack.
(6) Air purge inlet
Inlet for feeding instrumentation air. It is required to prevent lens conden-
sation and eliminate attached dust. Prepare the pipe to connect to joint for
tube
10/8. (Use the air for air purge which does not contain oil or mist.
Othewise, the oil builtup inside the flowmeter may reduce the flow or oil
attached to the lens surface may cause difficulty in light receiring.)
(7) Companion flange
Connects to the companion flange which is mounted on the stack. (This
flange should be prepared by users.)
(8) Cable between receiver unit and
transmitter unit
Supplies power to transmitter unit, and sends and receives signal between
receiver unit and transmitter unit. Up to 25m can be specified.
(9) Cable between receiver unit and
control unit
Supplies power to receiver unit, and sends and receives signal between
receiver unit and transmitter unit. Up to 100m can be specified.
(10) Control unit fixed fixture
Fixes control unit on the wall. There are two
12 holes as mounting holes
at upper part, and 12mm width U-tube hole at lower part.
(11) Display part
Displays the measurement value and alarm.
(12) External input/output part
Input/output part for AC power supply, signal cable to the receiver unit,
analog input/output and relay input/output. The hole diameter is
10mm.