- 54 -
INZ-TN7ZSS-E
6.2.5 Matching a value obtained by manual dust analysis
It may take several days until the result of manual analysis is obtained. After the result of manual analysis is
obtained, the concentration of dust average value measured while sampling gas for manual analysis was be-
ing extracted should be matched.
Point the cursor to [Component] on the
“Span Calibration” screen to select “Dust”,
and press the
key.
A value then can be entered to [Meas.]. En-
ter the average value measured by “Average.
Dust Value” screen.
Enter the value obtained from manual analy-
sis to Cal. Gas Value.
2006-04-01
12:10:00
Span Cal.Start
Cal.Finish
500
mm
1000.00
vol%
0025
°C
kPa
00.00
100.00
vol%
Range 1
2006/09/01
Dust
Component
Cal. Range
Cell Length
Cal.Gas Value
Meas.
Gas Temp.
Gas Press.
LastCal.Day
Span Calibration
ENT
ENT
-5+
2006-04-01
12:10:00
Span Cal.Start
Cal.Finish
500
mm
1000.00
vol%
0025
°C
kPa
00.00
100.00
vol%
Range 1
2006/09/01
Dust
Ave. Dust Value
Component
Cal. Range
Cell Length
Cal.Gas Value
Meas.
Gas Temp.
Gas Press.
LastCal.Day
Span Calibration
ENT
-5+
2006-04-01
12:10:00
Span Cal.Start
Cal.Finish
500
mm
1000.00
vol%
0025
°C
kPa
00.00
100.00
vol%
Range 1
2006/09/01
Dust
Ave. Dust Value
Component
Cal. Range
Cell Length
Cal.Gas Value
Meas.
Gas Temp.
Gas Press.
LastCal.Day
Span Calibration