![Fresenius Medical Care 5008 Operating Instructions Manual Download Page 72](http://html.mh-extra.com/html/fresenius-medical-care/5008/5008_operating-instructions-manual_2331649072.webp)
Chapter 5: Preparation
5-22
Fresenius Medical Care 5008 OP 5/09.06
Caution
When inserting the tubing systems, the following precautions
must be respected:
– The tubing systems have to be free of kinks, tension and twists and
must not be jammed (risk of hemolysis). Use the line holders
provided.
– Ensure the correct position of the screwed connections, especially
of the connection sites to the patient, the dialyzer and the system
and check or correct them during the treatment if necessary. Take
the appropriate measures if required (e.g. retightening of the Luer
Lock connection or replacement of the tubing system).
– Check the protective caps for tight fit and tighten them if necessary.
– The lines for the supply of infusions should always be clamped,
except if they are needed.
– During long-term operation, the blood lines must be changed after
24 hours at the latest.
– Do not use cannulas with a diameter of > 20 gauge to pierce the
septum of the injection sites. Insert the cannula vertically and in the
center of the septum. Disinfect the injection sites with 70% alcohol
before use.
– The blood pump must be set to the diameter of the pump segment,
refer also to the product label of the blood lines. If a wrong line
diameter is set, this may cause significant deviations in the blood
flow and thus in the dialysis dose.
– Materials which come directly or indirectly into contact with blood
are: Plasticized PVC, unplasticized PVC, polyethylene,
polycarbonate, latex-free rubber, ABS.
– The minimum temperature of the tubing systems during use is 18 °C.
Caution
Delivery operation of the pump(s) with open doors
(blood pump, substituate pump, optional Single-Needle pump):
When the doors are open and the rotor of the pump(s) is running, make
sure that
no objects
, such as fingers, hair or ball point pens, come into
contact with the rotor (risk of injury).
Arterial pressure measurement unit:
Prevent foreign objects from coming into contact with the arterial
pressure measurement unit.
Heparin pump:
If heparin syringes of third party suppliers are used, the operator is
responsible and has to ensure that the syringe data displayed match the
actual data.
Heparin syringes without Luer lock are not recommended as the
connection between the heparin syringe and the blood lines may come
loose. If heparin syringes without Luer lock are used it is the operator’s
responsibility to ensure that the connection between the heparin syringe
and the blood lines does not loosen inadvertently.
IV pole:
Securely fix bags or other objects to be hung from the IV pole.
Summary of Contents for 5008
Page 2: ......
Page 14: ...Chapter 1 Index 1 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 38: ...Chapter 3 Design 3 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 48: ...Chapter 4 Graphical User Interface 4 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 49: ...Fold Out Sheet 5 Preparation Fresenius Medical Care 5008 OP 5 09 06 ...
Page 66: ...Chapter 5 Preparation 5 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 96: ...Chapter 5 Preparation 5 46 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 111: ...Fold Out Sheet 6 Treatment Fresenius Medical Care 5008 OP 5 09 06 ...
Page 112: ...TREATMENT SCREEN SN setting the level in the SN chamber ...
Page 120: ...Chapter 6 Treatment 6 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 128: ...Chapter 6 Treatment 6 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 135: ...Fold Out Sheet 7 Reinfusion Fresenius Medical Care 5008 OP 5 09 06 ...
Page 136: ...REINFUSION SCREEN ...
Page 140: ...Chapter 7 Reinfusion 7 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 144: ...Chapter 7 Reinfusion 7 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 146: ...Chapter 7 Reinfusion 7 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 147: ...Fold Out Sheet 8 Cleaning Fresenius Medical Care 5008 OP 5 09 06 ...
Page 148: ...Connecting the disinfectant container CLEANING SCREEN Program run graphic e g Diasteril ...
Page 155: ...Fold Out Sheet 9 Alarm processing Fresenius Medical Care 5008 OP 5 09 06 ...
Page 175: ...Fold Out Sheet 10 Other Functions Fresenius Medical Care 5008 OP 5 09 06 ...
Page 210: ...Chapter 10 Other Functions 10 34 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 231: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 21 ...
Page 272: ...Chapter 13 Certificates 13 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 273: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 3 ...
Page 274: ...Chapter 13 Certificates 13 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 276: ...Chapter 14 Appendix 14 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 277: ...Fold Out Sheet 15 OCM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 278: ...OCM SCREEN OCM diagram ...
Page 284: ...Chapter 15 Option OCM Online Clearance Monitoring 15 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 285: ...Fold Out Sheet 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 ...
Page 286: ...ONLINE SCREEN ...
Page 287: ...Chapter 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 16 1 16 ONLINEplus ...
Page 290: ...Chapter 16 ONLINEplus 16 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 291: ...Fold Out Sheet 17 Option SN Fresenius Medical Care 5008 OP 5 09 06 ...
Page 292: ...SINGLE NEEDLE SCREEN ...
Page 296: ...Chapter 17 Option SN Single Needle 17 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 297: ...Fold Out Sheet 18 Option BPM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 304: ...Chapter 18 Option BPM Blood Pressure Monitoring 18 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 306: ...Chapter 19 Option CBPM 19 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 307: ...Fold Out Sheet 20 BTM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 313: ...Fold Out Sheet 21 BVM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 326: ...Chapter 23 Options BLK WET 23 4 Fresenius Medical Care 5008 OP 5 09 06 ...