Chapter 6: Treatment
6-18
Fresenius Medical Care 5008 OP 5/09.06
Message:
Blood detected - Dialysis
Start
The blood pump stops. The arterial/venous occlusion clamp closes.
Mute
LED is flashing.
Touch the
Start
button.
The system automatically changes to the TREATMENT SCREEN.
Heparin pump and ultrafiltration are started automatically, if preset in
the Operator Setup. Set the blood flow to the desired delivery rate. The
alarm limits will be set automatically.
Standard Connect
Connect the arterial line to the patient.
If infusion of the rinse solution contained in the line is desired, now
connect the venous line.
Touch the
Standard
Connect
button.
Message:
Start the blood pump –
Confirm
Touch the
Confirm
button.
The blood pump is delivering.
Set the delivery rate of the blood pump to the desired value.
Optical detector sensed blood.
Message:
Blood detected - Double-Needle dialysis
Start
– Single-
Needle dialysis
Start
The blood pump stops. The arterial/venous occlusion clamp closes.
Mute
LED is flashing.
Connect the venous connector of the tubing system to the patient
(provided this has not been done yet).
Touch the
Double-Needle dialysis
Start
button.
Caution
When using high-flux dialyzers and selecting low UF rates there is a
possibility of local backfiltration.
Backfiltration depends on:
– the type of high-flux dialyzer
– the different flow resistances on the dialysate and the blood side
– the viscosity of the blood
Note
If the level in the SN chamber rises too high during the treatment it has
to be reduced (see chapter 6.4.3 , page 6-19)
Caution
When using high-flux dialyzers and selecting low UF rates there is a
possibility of local backfiltration.
Backfiltration depends on:
– the type of high-flux dialyzer
– the different flow resistances on the dialysate and the blood side
– the viscosity of the blood
Summary of Contents for 5008
Page 2: ......
Page 14: ...Chapter 1 Index 1 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 38: ...Chapter 3 Design 3 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 48: ...Chapter 4 Graphical User Interface 4 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 49: ...Fold Out Sheet 5 Preparation Fresenius Medical Care 5008 OP 5 09 06 ...
Page 66: ...Chapter 5 Preparation 5 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 96: ...Chapter 5 Preparation 5 46 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 111: ...Fold Out Sheet 6 Treatment Fresenius Medical Care 5008 OP 5 09 06 ...
Page 112: ...TREATMENT SCREEN SN setting the level in the SN chamber ...
Page 120: ...Chapter 6 Treatment 6 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 128: ...Chapter 6 Treatment 6 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 135: ...Fold Out Sheet 7 Reinfusion Fresenius Medical Care 5008 OP 5 09 06 ...
Page 136: ...REINFUSION SCREEN ...
Page 140: ...Chapter 7 Reinfusion 7 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 144: ...Chapter 7 Reinfusion 7 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 146: ...Chapter 7 Reinfusion 7 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 147: ...Fold Out Sheet 8 Cleaning Fresenius Medical Care 5008 OP 5 09 06 ...
Page 148: ...Connecting the disinfectant container CLEANING SCREEN Program run graphic e g Diasteril ...
Page 155: ...Fold Out Sheet 9 Alarm processing Fresenius Medical Care 5008 OP 5 09 06 ...
Page 175: ...Fold Out Sheet 10 Other Functions Fresenius Medical Care 5008 OP 5 09 06 ...
Page 210: ...Chapter 10 Other Functions 10 34 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 231: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 21 ...
Page 272: ...Chapter 13 Certificates 13 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 273: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 3 ...
Page 274: ...Chapter 13 Certificates 13 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 276: ...Chapter 14 Appendix 14 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 277: ...Fold Out Sheet 15 OCM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 278: ...OCM SCREEN OCM diagram ...
Page 284: ...Chapter 15 Option OCM Online Clearance Monitoring 15 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 285: ...Fold Out Sheet 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 ...
Page 286: ...ONLINE SCREEN ...
Page 287: ...Chapter 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 16 1 16 ONLINEplus ...
Page 290: ...Chapter 16 ONLINEplus 16 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 291: ...Fold Out Sheet 17 Option SN Fresenius Medical Care 5008 OP 5 09 06 ...
Page 292: ...SINGLE NEEDLE SCREEN ...
Page 296: ...Chapter 17 Option SN Single Needle 17 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 297: ...Fold Out Sheet 18 Option BPM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 304: ...Chapter 18 Option BPM Blood Pressure Monitoring 18 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 306: ...Chapter 19 Option CBPM 19 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 307: ...Fold Out Sheet 20 BTM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 313: ...Fold Out Sheet 21 BVM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 326: ...Chapter 23 Options BLK WET 23 4 Fresenius Medical Care 5008 OP 5 09 06 ...