Chapter 11: System Description
Fresenius Medical Care 5008 OP 5/09.06
11-23
Arterial path
The blood passes from the patient's vascular access to the arterial
occlusion clamp. The heparin pump is located between the arterial line
clamp and the arterial pressure measurement unit, before the blood
pump. The heparin pump is used to administer heparin doses to the
blood at an adjustable rate. The arterial pressure measurement unit
measures and monitors the pressure caused by the blood pump on the
access. The pressure is transferred by a pressure dome which is
located in the tubing system and which is forced against the membrane
of the pressure measurement unit. After it has passed the arterial
pressure measurement unit, the blood reaches the blood pump. The
blood pump pumps the blood to the dialyzer.
Venous path
The venous pressure port is situated next to the dialyzer. where the
venous return pressure is measured and monitored within tight limits.
The venous bubble catcher is placed into the level detector. If the level
in the venous bubble catcher falls it will be detected by the level
detector. Below the venous bubble catcher the blood passes through
the optical detector (OD). This detector differentiates between OD light
(rinse solution or air in the tubing system) or OD dark (blood in the
tubing system). The OD is followed by the air bubble detector, which
prevents infusion of air. The blood then flows from the air bubble
detector via the venous occlusion clamp back to the patient.
Single-Needle Click-Clack
The blood is transported through the extracorporeal blood circuit in a
recurrent cycle. The cycle consists of the arterial and the venous
phases.
During the arterial phase, the arterial occlusion clamp is open, and the
venous occlusion clamp is closed. The blood pump is delivering. This
causes a pressure to be built up in the blood lines. When the venous
pressure has reached the upper SN pressure, the system switches to
the venous phase.
During the venous phase, the arterial occlusion clamp is closed, and the
venous occlusion clamp is open. The blood pump is not delivering. The
pressure built-up in the blood lines will be reduced. When the venous
pressure has reached the lower SN pressure, the system switches to
the arterial phase.
Blood alarms
A blood alarm occurs:
– If the arterial pressure is above or below the preset limits.
– If the venous pressure is above or below the preset limits.
– If the TMP is above or below the alarm limits.
– If the air bubble detector detects air.
– If the blood pump stops for
≥
60 seconds.
– In the event of a blood leak of the dialyzer.
– Level monitored in the venous bubble catcher (warning).
– During Single-Needle Click-Clack: cycle exceeded
If a blood alarm is initiated:
– The blood pump will stop.
– The substituate pump will stop (only with ONLINE
plus
™).
– the venous occlusion clamp is closed.
– A visual and an audible alarm will be given.
– Ultrafiltration will be stopped.
– The heparin pump is stopped.
Summary of Contents for 5008
Page 2: ......
Page 14: ...Chapter 1 Index 1 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 38: ...Chapter 3 Design 3 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 48: ...Chapter 4 Graphical User Interface 4 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 49: ...Fold Out Sheet 5 Preparation Fresenius Medical Care 5008 OP 5 09 06 ...
Page 66: ...Chapter 5 Preparation 5 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 96: ...Chapter 5 Preparation 5 46 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 111: ...Fold Out Sheet 6 Treatment Fresenius Medical Care 5008 OP 5 09 06 ...
Page 112: ...TREATMENT SCREEN SN setting the level in the SN chamber ...
Page 120: ...Chapter 6 Treatment 6 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 128: ...Chapter 6 Treatment 6 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 135: ...Fold Out Sheet 7 Reinfusion Fresenius Medical Care 5008 OP 5 09 06 ...
Page 136: ...REINFUSION SCREEN ...
Page 140: ...Chapter 7 Reinfusion 7 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 144: ...Chapter 7 Reinfusion 7 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 146: ...Chapter 7 Reinfusion 7 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 147: ...Fold Out Sheet 8 Cleaning Fresenius Medical Care 5008 OP 5 09 06 ...
Page 148: ...Connecting the disinfectant container CLEANING SCREEN Program run graphic e g Diasteril ...
Page 155: ...Fold Out Sheet 9 Alarm processing Fresenius Medical Care 5008 OP 5 09 06 ...
Page 175: ...Fold Out Sheet 10 Other Functions Fresenius Medical Care 5008 OP 5 09 06 ...
Page 210: ...Chapter 10 Other Functions 10 34 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 231: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 21 ...
Page 272: ...Chapter 13 Certificates 13 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 273: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 3 ...
Page 274: ...Chapter 13 Certificates 13 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 276: ...Chapter 14 Appendix 14 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 277: ...Fold Out Sheet 15 OCM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 278: ...OCM SCREEN OCM diagram ...
Page 284: ...Chapter 15 Option OCM Online Clearance Monitoring 15 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 285: ...Fold Out Sheet 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 ...
Page 286: ...ONLINE SCREEN ...
Page 287: ...Chapter 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 16 1 16 ONLINEplus ...
Page 290: ...Chapter 16 ONLINEplus 16 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 291: ...Fold Out Sheet 17 Option SN Fresenius Medical Care 5008 OP 5 09 06 ...
Page 292: ...SINGLE NEEDLE SCREEN ...
Page 296: ...Chapter 17 Option SN Single Needle 17 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 297: ...Fold Out Sheet 18 Option BPM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 304: ...Chapter 18 Option BPM Blood Pressure Monitoring 18 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 306: ...Chapter 19 Option CBPM 19 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 307: ...Fold Out Sheet 20 BTM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 313: ...Fold Out Sheet 21 BVM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 326: ...Chapter 23 Options BLK WET 23 4 Fresenius Medical Care 5008 OP 5 09 06 ...