Chapter 11: System Description
Fresenius Medical Care 5008 OP 5/09.06
11-9
UFC measurement
At the beginning of the treatment, a measurement of the ultrafiltration
coefficient (UFC) of the dialyzer connected is performed. This value is
taken as initial value for different parameters (e.g. TMP monitoring).
If this measurement can not be performed successfully 3 times because
of malfunctions, a message will be displayed:
3
rd
invalid UFC measure-
ment
with the appropriate information about possible causes. As long
as no current value is measured, a default value is used for the calcula-
tion instead of the measured value
.
Balancing
Accuracy: ±0.1 % related to the total dialysate volume
Maximum balancing error
Example:
Ultrafiltration error: with 1000 ml in 1 hour: ±1 % = ±10 ml/h
Balancing error: with 30 l fluid flow in 1 hour at a dialysate flow of
500 ml/min: ±0.1 %= ± 30 ml/h
Maximum balancing error:
F = F
UF
+ F
Bil
= (±10 ml/h) + (±30 ml/h) = ±40 ml/h
Degassing
Method: Negative pressure
Dialysate concentration
(conductivity)
Display range: 12.8 to 15.7 mS/cm
Resolution: 0.1 mS/cm
Accuracy: 0.1 mS/cm
Method:
Temperature-compensated electronic conductivity meter with
adjustable alarm limits.
Concentrates
Entering concentration types
Adjustment range: 125 to 151 mmol/l, depending on the concentrate
used ±10 % of the base value.
Bicarbonate readjustment range: corresponds to ±8 mmol/l
bi
b
ag
®
Bicarbonate concentrate preparation from the bi
b
ag
®
Temperature range: 15 to 35 °C
Dialysate temperature
Adjustment range: (prescribed temperature) 34.0 °C to 39.0 °C
Resolution: 0.5 °C
Measuring accuracy: + 0.2 °C / – 0.5 °C
(measuring accuracy under calibration conditions for a dialysate flow of 500 ml/min.)
Dialysate flow
Display range: 100 to 1000 ml/min
Resolution: 100 ml/min
Desired values: 100 to 1000 ml/min
Measurement by means of time pulse monitoring and balancing
chamber volume
AutoFlow: The AutoFlow function automatically regulates the dialysate
flow, depending on the dialyzer type and blood flow.
F
=
F
UF
+
F
Bil
F
=
Maximum balancing error
F
UF
=
Ultrafiltration error
F
Bil
=
Balancing error
Summary of Contents for 5008
Page 2: ......
Page 14: ...Chapter 1 Index 1 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 38: ...Chapter 3 Design 3 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 48: ...Chapter 4 Graphical User Interface 4 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 49: ...Fold Out Sheet 5 Preparation Fresenius Medical Care 5008 OP 5 09 06 ...
Page 66: ...Chapter 5 Preparation 5 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 96: ...Chapter 5 Preparation 5 46 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 111: ...Fold Out Sheet 6 Treatment Fresenius Medical Care 5008 OP 5 09 06 ...
Page 112: ...TREATMENT SCREEN SN setting the level in the SN chamber ...
Page 120: ...Chapter 6 Treatment 6 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 128: ...Chapter 6 Treatment 6 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 135: ...Fold Out Sheet 7 Reinfusion Fresenius Medical Care 5008 OP 5 09 06 ...
Page 136: ...REINFUSION SCREEN ...
Page 140: ...Chapter 7 Reinfusion 7 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 144: ...Chapter 7 Reinfusion 7 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 146: ...Chapter 7 Reinfusion 7 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 147: ...Fold Out Sheet 8 Cleaning Fresenius Medical Care 5008 OP 5 09 06 ...
Page 148: ...Connecting the disinfectant container CLEANING SCREEN Program run graphic e g Diasteril ...
Page 155: ...Fold Out Sheet 9 Alarm processing Fresenius Medical Care 5008 OP 5 09 06 ...
Page 175: ...Fold Out Sheet 10 Other Functions Fresenius Medical Care 5008 OP 5 09 06 ...
Page 210: ...Chapter 10 Other Functions 10 34 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 231: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 21 ...
Page 272: ...Chapter 13 Certificates 13 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 273: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 3 ...
Page 274: ...Chapter 13 Certificates 13 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 276: ...Chapter 14 Appendix 14 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 277: ...Fold Out Sheet 15 OCM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 278: ...OCM SCREEN OCM diagram ...
Page 284: ...Chapter 15 Option OCM Online Clearance Monitoring 15 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 285: ...Fold Out Sheet 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 ...
Page 286: ...ONLINE SCREEN ...
Page 287: ...Chapter 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 16 1 16 ONLINEplus ...
Page 290: ...Chapter 16 ONLINEplus 16 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 291: ...Fold Out Sheet 17 Option SN Fresenius Medical Care 5008 OP 5 09 06 ...
Page 292: ...SINGLE NEEDLE SCREEN ...
Page 296: ...Chapter 17 Option SN Single Needle 17 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 297: ...Fold Out Sheet 18 Option BPM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 304: ...Chapter 18 Option BPM Blood Pressure Monitoring 18 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 306: ...Chapter 19 Option CBPM 19 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 307: ...Fold Out Sheet 20 BTM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 313: ...Fold Out Sheet 21 BVM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 326: ...Chapter 23 Options BLK WET 23 4 Fresenius Medical Care 5008 OP 5 09 06 ...