Chapter 21: Option BVM (Blood Volume Monitor)
21-6
Fresenius Medical Care 5008 OP 5/09.06
Message
Message:
BVM: Achieving UF goal is uncertain. Reduce minimum UF
goal by XXX ml or set a minimum BVM rate of XXXX ml/h. – Minimum
goal
Reduce
– Minimum rate
Set
This message is displayed only during the second half of the volume to
be removed.
Touch the
Reduce
button to modify the prescription. The control
continues in the usual manner.
(If a PatientCard is used, this new minimum UF goal will also be used
for the next treatment.)
Touch the
Set
button to set the minimum BVM rate. The message is
displayed only once during the second half of the volume to be
removed. Until the end of the treatment, at least the minimum BVM rate
will be used (lower limitation of the BVM rate), even if the RBV continues
to decrease. If the patient recovers, treatment will be performed with a
higher BVM rate, according to the RBV.
The minimum UF goal prescribed will imperatively be respected.
After the UF control parameter (e.g. UF goal) was changed, the BVM
minimum rate is automatically cleared.
Message
Message:
BVM: Clotting in dialyzer possible. Blood flow too low
compared to the minimum BVM rate. Increase the blood flow to a
minimum of XXX ml/min.
–
Confirm
Touch the
Confirm
button.
If possible, increase the blood flow at least up to the indicated value. If
not, reduce the UF goal or prolong the UF time.
Summary of Contents for 5008
Page 2: ......
Page 14: ...Chapter 1 Index 1 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 38: ...Chapter 3 Design 3 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 48: ...Chapter 4 Graphical User Interface 4 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 49: ...Fold Out Sheet 5 Preparation Fresenius Medical Care 5008 OP 5 09 06 ...
Page 66: ...Chapter 5 Preparation 5 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 96: ...Chapter 5 Preparation 5 46 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 111: ...Fold Out Sheet 6 Treatment Fresenius Medical Care 5008 OP 5 09 06 ...
Page 112: ...TREATMENT SCREEN SN setting the level in the SN chamber ...
Page 120: ...Chapter 6 Treatment 6 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 128: ...Chapter 6 Treatment 6 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 135: ...Fold Out Sheet 7 Reinfusion Fresenius Medical Care 5008 OP 5 09 06 ...
Page 136: ...REINFUSION SCREEN ...
Page 140: ...Chapter 7 Reinfusion 7 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 144: ...Chapter 7 Reinfusion 7 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 146: ...Chapter 7 Reinfusion 7 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 147: ...Fold Out Sheet 8 Cleaning Fresenius Medical Care 5008 OP 5 09 06 ...
Page 148: ...Connecting the disinfectant container CLEANING SCREEN Program run graphic e g Diasteril ...
Page 155: ...Fold Out Sheet 9 Alarm processing Fresenius Medical Care 5008 OP 5 09 06 ...
Page 175: ...Fold Out Sheet 10 Other Functions Fresenius Medical Care 5008 OP 5 09 06 ...
Page 210: ...Chapter 10 Other Functions 10 34 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 231: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 21 ...
Page 272: ...Chapter 13 Certificates 13 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 273: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 3 ...
Page 274: ...Chapter 13 Certificates 13 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 276: ...Chapter 14 Appendix 14 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 277: ...Fold Out Sheet 15 OCM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 278: ...OCM SCREEN OCM diagram ...
Page 284: ...Chapter 15 Option OCM Online Clearance Monitoring 15 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 285: ...Fold Out Sheet 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 ...
Page 286: ...ONLINE SCREEN ...
Page 287: ...Chapter 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 16 1 16 ONLINEplus ...
Page 290: ...Chapter 16 ONLINEplus 16 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 291: ...Fold Out Sheet 17 Option SN Fresenius Medical Care 5008 OP 5 09 06 ...
Page 292: ...SINGLE NEEDLE SCREEN ...
Page 296: ...Chapter 17 Option SN Single Needle 17 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 297: ...Fold Out Sheet 18 Option BPM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 304: ...Chapter 18 Option BPM Blood Pressure Monitoring 18 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 306: ...Chapter 19 Option CBPM 19 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 307: ...Fold Out Sheet 20 BTM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 313: ...Fold Out Sheet 21 BVM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 326: ...Chapter 23 Options BLK WET 23 4 Fresenius Medical Care 5008 OP 5 09 06 ...