![Fresenius Medical Care 5008 Operating Instructions Manual Download Page 221](http://html.mh-extra.com/html/fresenius-medical-care/5008/5008_operating-instructions-manual_2331649221.webp)
Chapter 11: System Description
Fresenius Medical Care 5008 OP 5/09.06
11-11
Flow alarm
Dependent on the programmed flow
11.1.12 Extracorporeal Blood Circuit and Safety Systems
Arterial pressure
measurement
Display range: –300 to +300 mmHg
Resolution: 5 mmHg
Accuracy: 7 mmHg (typical)
OD senses light:
Alarm window width: –300 to +300 mmHg
OD senses dark:
Alarm window width: +40 to +200 mmHg
Default value adjustable in the Operator setup, factory setting
120 mmHg
Blood pump
Delivery rate: 30 to 600 ml/min
Resolution: 10 ml/min (with a line diameter of 8 mm)
Accuracy: ±10 %
Line diameter: 4.4 mm, 6.4 mm, 8.0 mm
Blood pump stop alarm: 60 seconds
(During Single-Needle operation - option - 180 seconds.)
Spring-loaded rollers, fully occluding, pressure-limited to 2 bar with 8 x
2.1 pump line segment (when using the prescribed tubing systems).
Venous pressure
measurement
Display range: –100 to +500 mmHg
Resolution: 5 mmHg
Accuracy: 7 mmHg (typical)
OD senses light:
Alarm window width: –100 to +500 mmHg
OD senses dark:
Alarm window width: 40 to 200 mmHg
Default value adjustable in the Operator setup,
Factory setting 120 mmHg
adjustable over a range of 20 to 500 mmHg
(adjustable in the Operator setup from -100 to 500 mmHg.)
Level detector
Method:
Capacitive measurement
Switching point 13 mm, ±4 mm from upper edge
Optical detector
Method: Infrared transmission
Distinguishes between
OD light (rinse solution or air in the tubing system)
OD dark (blood in the tubing system).
Air bubble detector
Method:
Ultrasonic transmission measurement on the line
Sensitivity:
– Air bubbles: Bubble volume
≥
20 µl
– Blood foam (air-blood mixture)
– Micro bubbles
Summary of Contents for 5008
Page 2: ......
Page 14: ...Chapter 1 Index 1 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 38: ...Chapter 3 Design 3 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 48: ...Chapter 4 Graphical User Interface 4 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 49: ...Fold Out Sheet 5 Preparation Fresenius Medical Care 5008 OP 5 09 06 ...
Page 66: ...Chapter 5 Preparation 5 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 96: ...Chapter 5 Preparation 5 46 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 111: ...Fold Out Sheet 6 Treatment Fresenius Medical Care 5008 OP 5 09 06 ...
Page 112: ...TREATMENT SCREEN SN setting the level in the SN chamber ...
Page 120: ...Chapter 6 Treatment 6 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 128: ...Chapter 6 Treatment 6 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 135: ...Fold Out Sheet 7 Reinfusion Fresenius Medical Care 5008 OP 5 09 06 ...
Page 136: ...REINFUSION SCREEN ...
Page 140: ...Chapter 7 Reinfusion 7 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 144: ...Chapter 7 Reinfusion 7 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 146: ...Chapter 7 Reinfusion 7 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 147: ...Fold Out Sheet 8 Cleaning Fresenius Medical Care 5008 OP 5 09 06 ...
Page 148: ...Connecting the disinfectant container CLEANING SCREEN Program run graphic e g Diasteril ...
Page 155: ...Fold Out Sheet 9 Alarm processing Fresenius Medical Care 5008 OP 5 09 06 ...
Page 175: ...Fold Out Sheet 10 Other Functions Fresenius Medical Care 5008 OP 5 09 06 ...
Page 210: ...Chapter 10 Other Functions 10 34 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 231: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 21 ...
Page 272: ...Chapter 13 Certificates 13 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 273: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 3 ...
Page 274: ...Chapter 13 Certificates 13 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 276: ...Chapter 14 Appendix 14 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 277: ...Fold Out Sheet 15 OCM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 278: ...OCM SCREEN OCM diagram ...
Page 284: ...Chapter 15 Option OCM Online Clearance Monitoring 15 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 285: ...Fold Out Sheet 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 ...
Page 286: ...ONLINE SCREEN ...
Page 287: ...Chapter 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 16 1 16 ONLINEplus ...
Page 290: ...Chapter 16 ONLINEplus 16 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 291: ...Fold Out Sheet 17 Option SN Fresenius Medical Care 5008 OP 5 09 06 ...
Page 292: ...SINGLE NEEDLE SCREEN ...
Page 296: ...Chapter 17 Option SN Single Needle 17 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 297: ...Fold Out Sheet 18 Option BPM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 304: ...Chapter 18 Option BPM Blood Pressure Monitoring 18 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 306: ...Chapter 19 Option CBPM 19 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 307: ...Fold Out Sheet 20 BTM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 313: ...Fold Out Sheet 21 BVM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 326: ...Chapter 23 Options BLK WET 23 4 Fresenius Medical Care 5008 OP 5 09 06 ...