Chapter 11: System Description
11-10
Fresenius Medical Care 5008 OP 5/09.06
Entering the factor (AutoFlow) will modify the ratio of selected blood flow
to dialysate flow. The default value of the factor (AutoFlow) in the
Operator setup is 1.5.
Example:
Blood flow: 300 ml/min, factor: 1.5
Dialysate flow = 450 ml/min
The factor (AutoFlow) may be modified in the Operator setup and in the
DIALYSATE SCREEN.
Values above 1.5 may result in a minor increase of the dialysis dose. At
the same time, the water, concentrate and energy consumption
increases.
The contrary applies to values below 1.5, i.e., the dialysis dose may
decrease, while the water, concentrate and energy consumption may
be reduced.
EcoFlow: dialysate flow automatically reduced to 100 ml/min in
Preparation
The following has to be observed regarding the dialysate flow:
If the water inlet rate is not sufficient for achieving the maximum
dialysate flow of 1000 ml/min, the admissible adjustment range will be
delimited accordingly.
Rinse temperature
Rinsing:
Desired temperature: 37 °C
Resolution: 0.5 °C
Measuring accuracy: ±0.2 °C
Hot rinse:
Desired temperature: 85 °C
Resolution: 0.5 °C
Measuring accuracy: ±2.0 °C
Disinfection temperature
Cold disinfection:
Desired temperature: 37 °C
Resolution: 0.5 °C
Measuring accuracy: ±0.2 °C
Degreasing/cold disinfection:
Desired temperature: 37 °C
Resolution: 0.5 °C
Measuring accuracy: ±0.2 °C
Heat disinfection:
Desired temperature: 85 °C
Resolution: 0.5 °C
Measuring accuracy: ±2.0 °C
Rinse and disinfection flow
Desired value: 600 ml/min or 800 ml/min (depending on the settings in
the Technician's setup)
Concentration of
disinfectant
Dilution:
The disinfectants (Citrosteril
®
, Diasteril
®
, Puristeril
®
340,
Puristeril
®
plus
) are diluted with purified water in the dialysis system at
a ratio of 1+24.
The disinfectant (Sporotal
®
100) is diluted with purified water in the
dialysis system at a ratio of 1+34.
Summary of Contents for 5008
Page 2: ......
Page 14: ...Chapter 1 Index 1 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 38: ...Chapter 3 Design 3 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 48: ...Chapter 4 Graphical User Interface 4 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 49: ...Fold Out Sheet 5 Preparation Fresenius Medical Care 5008 OP 5 09 06 ...
Page 66: ...Chapter 5 Preparation 5 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 96: ...Chapter 5 Preparation 5 46 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 111: ...Fold Out Sheet 6 Treatment Fresenius Medical Care 5008 OP 5 09 06 ...
Page 112: ...TREATMENT SCREEN SN setting the level in the SN chamber ...
Page 120: ...Chapter 6 Treatment 6 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 128: ...Chapter 6 Treatment 6 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 135: ...Fold Out Sheet 7 Reinfusion Fresenius Medical Care 5008 OP 5 09 06 ...
Page 136: ...REINFUSION SCREEN ...
Page 140: ...Chapter 7 Reinfusion 7 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 144: ...Chapter 7 Reinfusion 7 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 146: ...Chapter 7 Reinfusion 7 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 147: ...Fold Out Sheet 8 Cleaning Fresenius Medical Care 5008 OP 5 09 06 ...
Page 148: ...Connecting the disinfectant container CLEANING SCREEN Program run graphic e g Diasteril ...
Page 155: ...Fold Out Sheet 9 Alarm processing Fresenius Medical Care 5008 OP 5 09 06 ...
Page 175: ...Fold Out Sheet 10 Other Functions Fresenius Medical Care 5008 OP 5 09 06 ...
Page 210: ...Chapter 10 Other Functions 10 34 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 231: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 21 ...
Page 272: ...Chapter 13 Certificates 13 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 273: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 3 ...
Page 274: ...Chapter 13 Certificates 13 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 276: ...Chapter 14 Appendix 14 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 277: ...Fold Out Sheet 15 OCM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 278: ...OCM SCREEN OCM diagram ...
Page 284: ...Chapter 15 Option OCM Online Clearance Monitoring 15 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 285: ...Fold Out Sheet 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 ...
Page 286: ...ONLINE SCREEN ...
Page 287: ...Chapter 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 16 1 16 ONLINEplus ...
Page 290: ...Chapter 16 ONLINEplus 16 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 291: ...Fold Out Sheet 17 Option SN Fresenius Medical Care 5008 OP 5 09 06 ...
Page 292: ...SINGLE NEEDLE SCREEN ...
Page 296: ...Chapter 17 Option SN Single Needle 17 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 297: ...Fold Out Sheet 18 Option BPM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 304: ...Chapter 18 Option BPM Blood Pressure Monitoring 18 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 306: ...Chapter 19 Option CBPM 19 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 307: ...Fold Out Sheet 20 BTM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 313: ...Fold Out Sheet 21 BVM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 326: ...Chapter 23 Options BLK WET 23 4 Fresenius Medical Care 5008 OP 5 09 06 ...