Chapter 11: System Description
Fresenius Medical Care 5008 OP 5/09.06
11-29
The ONLINE
plus
™ option consists primarily of a substituate pump, the
substituate port (to connect the SafeLine™ disposable as substituate
line), a rinse port (to rinse the dialyzer) and two DIASAFE
®
plus
filters.
Apart from the preparation of the substituate from the dialysate, the
ONLINE
plus
™ also provides the following functions:
– Priming/rinsing the extracorporeal blood circuit with substituate
– Infinite rinsing of the EBC and the dialyzer
– Bolus with substituate; a bolus can also be given in HD, provided the
SafeLine™ is connected (the bolus will NOT be balanced!)
– Reinfusion (blood return) with substituate
O
Intended use
Fields of application
In addition to hemodialysis (HD) the ONLINE
plus
™ option also
supports the following procedures (the treatment mode can be changed
during the treatment):
– Hemofiltration (HF) pre-/postdilution
– Hemodiafiltration (HDF) pre-/postdilution
– ONLINE priming and rinsing of the extracorporeal blood circuit
– Bolus administration in case of hypotension
Chemical composition of
the substitution fluid
produced online
The chemical composition of the fluid produced with the ONLINE
plus
™
is identical to the composition of the fresh dialysate selected by the
operator. This applies to all electrolytes contained in the dialysate and
especially the bicarbonate content of the fluid.
The amount of bicarbonate to be considered is determined by:
Restrictions
An HF treatment may not be performed with a dialyzer with one filtrate
connector (HF filter without dialysate couplings).
11.5.7 SN Description (Option)
During SN dialysis, blood is alternately collected from and returned to
the patient through one needle. (2-pump procedure).
When the arterial clamp is open the arterial blood pump rotates and
delivers blood into the SN chamber. The volume delivered depends on
the stroke volume set (10 to 50 ml). The Single-Needle pump is stopped
and the venous occlusion clamp is closed. The external compliance
chamber is connected to the Single-Needle pressure transducer. The
internal compliance chamber located within the EBM is filled with air.
The arterial blood pump stops and the arterial clamp closes as soon as
the stroke volume and pressure is achieved. The SN blood pump
rotates and pumps the blood through the dialyzer, the venous bubble
catcher and the venous occlusion clamp into the patient. As soon as the
pressure in the internal compliance chamber reduces below the lower
set limit, the process described above restarts.
Quantity
bicarbonate [mmol]
=
volume
infused [l]
X bicarbonate concentration of
the dialysate [mmol/l]
Summary of Contents for 5008
Page 2: ......
Page 14: ...Chapter 1 Index 1 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 38: ...Chapter 3 Design 3 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 48: ...Chapter 4 Graphical User Interface 4 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 49: ...Fold Out Sheet 5 Preparation Fresenius Medical Care 5008 OP 5 09 06 ...
Page 66: ...Chapter 5 Preparation 5 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 96: ...Chapter 5 Preparation 5 46 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 111: ...Fold Out Sheet 6 Treatment Fresenius Medical Care 5008 OP 5 09 06 ...
Page 112: ...TREATMENT SCREEN SN setting the level in the SN chamber ...
Page 120: ...Chapter 6 Treatment 6 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 128: ...Chapter 6 Treatment 6 16 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 135: ...Fold Out Sheet 7 Reinfusion Fresenius Medical Care 5008 OP 5 09 06 ...
Page 136: ...REINFUSION SCREEN ...
Page 140: ...Chapter 7 Reinfusion 7 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 144: ...Chapter 7 Reinfusion 7 8 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 146: ...Chapter 7 Reinfusion 7 10 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 147: ...Fold Out Sheet 8 Cleaning Fresenius Medical Care 5008 OP 5 09 06 ...
Page 148: ...Connecting the disinfectant container CLEANING SCREEN Program run graphic e g Diasteril ...
Page 155: ...Fold Out Sheet 9 Alarm processing Fresenius Medical Care 5008 OP 5 09 06 ...
Page 175: ...Fold Out Sheet 10 Other Functions Fresenius Medical Care 5008 OP 5 09 06 ...
Page 210: ...Chapter 10 Other Functions 10 34 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 231: ...Chapter 11 System Description Fresenius Medical Care 5008 OP 5 09 06 11 21 ...
Page 272: ...Chapter 13 Certificates 13 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 273: ...Chapter 13 Certificates Fresenius Medical Care 5008 OP 5 09 06 13 3 ...
Page 274: ...Chapter 13 Certificates 13 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 276: ...Chapter 14 Appendix 14 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 277: ...Fold Out Sheet 15 OCM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 278: ...OCM SCREEN OCM diagram ...
Page 284: ...Chapter 15 Option OCM Online Clearance Monitoring 15 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 285: ...Fold Out Sheet 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 ...
Page 286: ...ONLINE SCREEN ...
Page 287: ...Chapter 16 ONLINEplus Fresenius Medical Care 5008 OP 5 09 06 16 1 16 ONLINEplus ...
Page 290: ...Chapter 16 ONLINEplus 16 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 291: ...Fold Out Sheet 17 Option SN Fresenius Medical Care 5008 OP 5 09 06 ...
Page 292: ...SINGLE NEEDLE SCREEN ...
Page 296: ...Chapter 17 Option SN Single Needle 17 4 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 297: ...Fold Out Sheet 18 Option BPM Fresenius Medical Care 5008 OP 5 09 06 ...
Page 304: ...Chapter 18 Option BPM Blood Pressure Monitoring 18 6 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 306: ...Chapter 19 Option CBPM 19 2 Fresenius Medical Care 5008 OP 5 09 06 ...
Page 307: ...Fold Out Sheet 20 BTM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 313: ...Fold Out Sheet 21 BVM Option Fresenius Medical Care 5008 OP 5 09 06 ...
Page 326: ...Chapter 23 Options BLK WET 23 4 Fresenius Medical Care 5008 OP 5 09 06 ...