74
Calibration
2.2.
Set qty
shows the quantity of cleaning agent to be dispensed by the automatic feed-
er.
2.3.
Runtime
shows the time in which the automatic feeder should dispense the cleaning
agent.
2.4.
Date
shows when the cleaning agent was last calibrated.
2.5.
Confirm
Start?
by choosing
.
2.6.
Confirm
Exit automatic mode?
by choosing
. This message will only be dis-
played if your automatic feeder is still in automatic mode.
2.7.
Place the calibration box into the mixer beaker and confirm the prompt
Calibration
box used?
by choosing
.
Cleaning agent will be dispensed into the calibration beaker.
2.8.
The device now performs two check weighings without the user having to do anything
and shows the result.
2.9.
Then the average value of these two control weighings is shown flashing in the
dis-
played line
. Confirm with
. You then return to the calibration menu.
2.10.
Date
now shows the current date.
2.11. Remove the calibration beaker from the mixer beaker.
2.12. Repeat the calibration to check your results.
3. If the automatic feeder is equipped with a second cleaning agent pump, also calibrate the
Flushing agent 2
.
WARNING!
Risk of injury and death!
Never mix alkaline and acidic cleaning agents, since this may cause a dangerous chemical re-
action. Dangerous gases may be produced and cause serious breathing difficulties. They may
also cause explosions.
►
Empty the calibration beaker and rinse it out with clean water before using it to calibrate the
second cleaning agent.