Issue/Rev. 0.2 (12/12)
Page 98
•
MN0M023
7 Block OUTPUTS
Device Functions Proline Promass 83
Function description
OUTPUTS
o
PULSE/FREQUENCY OUTPUT (1...2)
o
OPERATION (PULSE)
SIMULATION
PULSE(4322)
!
Note!
This function is not available unless the PULSE option was selected in the OPERATION
MODE function.
Use this function to activate simulation of the pulse output.
Options:
OFF
COUNTDOWN
The pulses specified in the VALUE SIMULATION PULSE function are output.
CONTINUOUSLY
Pulses are continuously output with the pulse width specified in the PULSE WIDTH
function. Simulation is started once the CONTINUOUSLY option is confirmed with the
F
key.
!
Note!
Simulation is started by confirming the CONTINUOUSLY option with the
F
key. The
simulation can be switched off again via the SIMULATION PULSE function.
Factory setting:
OFF
!
Note!
• The notice message #631 “SIM. PULSE” indicates that simulation is active.
• The on/off ratio is 1:1 for both types of simulation.
• The measuring device continues to measure while simulation is in progress, i.e. the
current measured values are output correctly via the other outputs.
"
Caution!
The setting is not saved if the power supply fails.
VALUE SIMULATION
PULSE
(4323)
!
Note!
This function is not available unless the COUNTDOWN option was selected in the
SIMULATION PULSE function.
Use this function to specify the number of pulses (e.g. 50) which are output during the
simulation. This value is used to test downstream devices and the measuring device itself.
The pulses are output with the pulse width specified in the PULSE WIDTH function. The
on/off ratio is 1:1.
Simulation is started once the specified value is confirmed with the
F
key. The display
remains at 0 if the specified pulses have been output.
User input:
0...10 000
Factory setting:
0
!
Note!
Simulation is started by confirming the simulation value with the
F
key. The simulation
can be switched off again via the SIMULATION PULSE function.
"
Caution!
The setting is not saved if the power supply fails.