Issue/Rev. 0.2 (12/12)
Page 168
•
MN0M023
10 Block SPECIAL FUNCTION
Device Functions Proline Promass 83
10.3.3
Function group MASS FLOW
SPECIAL FUNCTION
H
ADVANCED DIAGNOSIS HEA
CONFIGURATION
740
ACQUISITION
741
MASS FLOW
742
Function description
SPECIAL FUNCTION
o
ADVANCED DIAGNOSIS
o
MASS FLOW
!
Note!
The appropriate unit is taken from the function UNIT MASS FLOW (0400) (see page 17).
REFERENCE VALUES
MASS FLOW
(7420)
The reference value for the mass flow appears on the display.
Display:
5-digit floating-point number, including unit and sign
MASS FLOW
(7421)
The measured mass flow appears on the display.
Display:
5-digit floating-point number, including unit and sign
MINIMUM VALUE
(7422)
The lowest mass flow value since the saved values were last reset appears on the display.
Display:
5-digit floating-point number, including unit and sign
MAXIMUM VALUE
(7423)
The highest mass flow value since the saved values were last reset appears on the display.
Display:
5-digit floating-point number, including unit and sign
MASS FLOW HISTORY
(7424)
The last ten mass flow values since the saved values were last reset appear on the display.
Display:
5-digit floating-point number, including unit and sign
MASS FLOW DEVIATION
(7425)
This function displays deviation between the measured mass flow and the reference
values (FACTORY or USER), see page 166, selected in the function SELECT REFERENCE
CONDITION (7402).
Display:
5-digit floating-point number, including unit and sign
WARNING LEVEL
(7426)
!
Note!
This function is not available unless ON was selected in the WARNING MODE function
(7403).
Use this function to specify a limit value for the mass flow. A notice message is generated
if the limit value is exceeded.
User input:
0...99999 [Mass flow unit]
Factory setting: