![FMC Technologies Proline Promass 83 E Description Of Device Functions Download Page 23](http://html.mh-extra.com/html/fmc-technologies/proline-promass-83-e/proline-promass-83-e_description-of-device-functions_2312629023.webp)
Issue/Rev. 0.2 (12/12)
MN0M023 • Page 23
Device Functions Proline Promass 83
3 Block MEASURED VARIABLES
TEXT ARBITRARY
VOLUME
(0602)
Use this function to enter a text for the selectable volume unit / volume flow unit. You
define only the text, the unit of time is provided from a choice of options (s, min, h, day).
User input:
xxxxxxx (max. 4 characters)
Valid characters are A-Z, 0-9, +,
-
, decimal point, white space or underscore
Factory setting:
“_ _ _ _” (No text)
Example:
If your text entry is “GLAS”, this text string appears on the display complete with the unit
of time, e.g. “GLAS/min”:
GLAS = Volume (text input)
GLAS / min = Volume flow as shown (on the display)
FACTOR ARBITRARY
VOLUME
(0603)
Use this function to define a quantity factor (without time) for the selectable unit. The
volume unit on which this factor is based is one liter.
User input:
7-digit floating-point number
Factory setting:
1
Reference quantity:
Liter
Example:
The volume of a glass is 0.5 l
o
2 glasses = 1 liter
User input: 2
TEXT ARBITRARY
DENSITY
(0604)
Use this function to enter a text for the selectable density unit.
User input:
xxxxxxx (max. 4 characters)
Valid characters are A-Z, 0-9, +,
-
, decimal point, white space or underscore
Factory setting:
“_ _ _ _” (No text)
Example:
Enter text “CE_L” (for centner per liter).
FACTOR ARBITRARY
DENSITY
(0605)
Use this function to define a quantity factor for the selectable density unit. The density
unit on which this factor is based is one kg/l.
User input:
7-digit floating-point number
Factory setting:
1
Reference quantity:
kg/l
Example:
One centner per liter is equivalent to 50 kg/l
o
0.02 centner/l = 1 kg/l
User input: 0.02
Function description
MEASURED VARIABLES
o
SPECIAL-UNITS
o
ARBITRARY UNIT