3 Block MEASURED VARIABLES
% CARRIER VOLUME
FLOW
(0028)
!
Note!
This function is not available unless one of the following was selected:
• in the function DENSITY FUNCTION (7000), see page 141:
– % MASS / % VOLUME
– FLEXIBLE and in the function MODE (7010), see page 144, the selection %
VOLUME 2D or % VOLUME 3D
Use this function to display the currently measured volume flow of the carrier fluid as a %
(of the overall mass flow). Carrier fluid = transporting liquid (e.g. water).
Display:
5-digit floating-point number, including unit and sign
CORRECTED CARRIER
VOLUME FLOW
(0029)
!
Note!
This function is not available unless % MASS / % VOLUME was selected in the DENSITY
FUNCTION function (7000), (see page 141).
Use this function to display the currently measured corrected volume flow of the carrier
fluid. Carrier fluid = transporting liquid (e.g. water).
Display:
5-digit floating-point number, including unit and sign
% BLACK LIQUOR
(0030)
!
Note!
This function is not available unless %-BLACK LIQUOR was selected in the DENSITY
FUNCTION function (7000), (see page 141).
The concentration in %-BLACK LIQUOR is displayed.
Display:
5-digit floating-point number, incl. units
°BAUME
(0031)
!
Note!
This function is not available unless °BAUME was selected in the DENSITY FUNCTION
function (7000), (see page 141).
The concentration in °BAUME is displayed.
Display:
5-digit floating-point number, incl. units
°API
(0033)
!
Note!
This function is not available unless °API was selected in the DENSITY FUNCTION
function (7000), (see page 141).
The concentration in °API is displayed.
Display:
5-digit floating-point number, incl. units