Issue/Rev. 0.2 (12/12)
Page 166
•
MN0M023
10 Block SPECIAL FUNCTION
Device Functions Proline Promass 83
10.3
Group ADVANCED DIAGNOSIS
10.3.1
Function group CONFIGURATION
SPECIAL FUNCTION
H
ADVANCED DIAGNOSIS HEA
CONFIGURATION
740
Function description
SPECIAL FUNCTION
o
ADVANCED DIAGNOSIS
o
CONFIGURATION
REFERENCE
CONDITION USER
(7401)
Use this function to start determining the user reference status. The following values are
determined:
• MASS FLOW
• DENSITY
• REFERENCE DENSITY
• TEMPERATURE
• TUBE DAMPING
• ELECTRODYNAMIC SENSORS
• OPERATING FREQUENCY FLUCTUATION
• TUBE DAMPING FLUCTUATION
Options:
CANCEL
START
Factory setting:
CANCEL
SELECT REFERENCE
CONDITION
(7402)
Use this function to select the reference status which should be used to compare the
advanced diagnosis parameters (see function ACQUISITION MODE (7410) on
page 167).
Options:
FACTORY
USER
Factory setting:
FACTORY
WARNING MODE
(7403)
Use this function to determine whether a warning should be generated when there is a
deviation between the reference status (FACTORY or USER, see function SELECT
REFERENCE CONDITION (7402)) and the current measuring values.
The values of the following functions are compared to the reference status :
• MASS FLOW (7421)
• DENSITY (7431)
• REFERENCE DENSITY (7441)
• TEMPERATURE (7451)
• TUBE DAMPING (7461)
• ELECTRODYNAMIC SENSORS (7471)
• OPERATING FREQUENCY FLUCTUATION (7481)
• TUBE DAMPING FLUCTUATION (7491)
Options:
OFF
ON
Factory setting:
OFF