Issue/Rev. 0.2 (12/12)
MN0M023 • Page 177
Device Functions Proline Promass 83 10 Block
SPECIAL
FUNCTION
10.3.10 Function group TUBE DAMPING FLUCTUATION
SPECIAL FUNCTION
H
ADVANCED DIAGNOSIS HEA
CONFIGURATION
740
ACQUISITION
741
MASS FLOW
742
DENSITY
743
REFERENCE DENSITY
744
TEMPERATURE
745
TUBE DAMPING
746
ELEC.-DYN. SENSORS
747
OPERAT. FREQ. FLUCT.
748
TUBE DAMP. FLUCT.
749
Function description
SPECIAL FUNCTION
o
ADVANCED DIAGNOSIS
o
TUBE DAMPING FLUCTUATION
REFERENCE VALUE
TUBE DAMPING
FLUCTUATION
(7490)
The reference value for the fluctuation of the tube damping appears on the display.
Display:
5-digit floating-point number
TUBE DAMPING
FLUCTUATION
(7491)
The measured fluctuation of the tube damping appears on the display.
Display:
5-digit floating-point number
MINIMUM VALUE
(7492)
The lowest value of the tube damping since the saved values were last reset appears on
the display.
Display:
5-digit floating-point number
MAXIMUM VALUE
(7493)
The highest value of the tube damping fluctuation since the saved values were last reset
appears on the display.
Display:
5-digit floating-point number
TUBE DAMPING
FLUCTUATION HISTORY
(7494)
The last ten values of the tube damping fluctuation since the saved values were last reset
appear on the display.
Display:
5-digit floating-point number
TUBE DAMPING
FLUCTUATION
DEVIATION
(7495)
This function displays the deviation between the measured deviation of the tube damping
and the reference values (FACTORY or USER) selected in the function SELECT
REFERENCE CONDITION (7402) see page 166.
Display:
5-digit floating-point number