9500B
Calibration Manual
64
Table 15. Sine Flatness Verification at 100 mV p-p into 50
Ω
Load
Verif
Point
Nominal
Output
Voltage
Output
Frequency
Measured
p-p Voltage
at 50 kHz
Specification
Relative to 50 kHz
User’s Total
Measurement
Uncertainty
Flatness
Acceptance
Limits
Measured
p-p
Voltage
for
Flatness
Check
(% of Output)
(Um)
Lower
Higher
Ref 4
100 mV
50 kHz
SF39
10
MHz
±2.0
SF40
50
MHz
±2.0
SF41
100
MHz
±2.0
SF42
250
MHz
±2.0
SF43
400
MHz
±2.5
SF44
550
MHz
±2.5
SF45
600
MHz
±3.5
SF46†
725
MHz
±3.5
SF47†
1
GHz
±3.5
SF48††
1.5
GHz
±4.0
SF49††
2
GHz
±4.0
SF50††
2.5
GHz
±4.0
SF51††
3
GHz
±4.0
SF68†††
4
GHz
±4.0
SF69†††
5
GHz
±4.0
SF70†††
5.5
GHz
±4.0
SF71†††
6
GHz
±4.0
Additional verification points for: 9500B/1100 and 9500B/3200 (†); 9500B/3200 only (††) and
9500B/3200 with 9560 (†††).