9500B
Calibration Manual
66
Interconnections
erw209
Figure 11. Edge Function Verification — Interconnections
Verification Setup
1.
Connections:
Ensure that the 9500B is connected to the
Oscilloscope as shown in Figure 11 and that both
instruments are powered on and warmed up.
2.
Oscilloscope:
Select the required function to measure edge
response.
W
Caution
The 9500B Edge function output is capable of generating
voltages that may cause damage to sampling oscilloscope
inputs with limited input voltage capability. Use of an attenuator
is typically required for outputs above 1 V pk-pk.
3.
9500B:
Ensure that the 9500B is in MANUAL mode and then
select the Edge function (
Ι
key). Select the required output
Signal Channel (50
Ω
Load), Trigger Channel, Cable
Select and Trigger Ratio.