9500B
Calibration Manual
132
Calibration Procedure: 150 ps Edge: Linearity
1.
Ensure that the 9500B is connected to the oscilloscope as shown in Figure 20.
and that both instruments are powered on and warmed up.
2.
Select the required 9500B Trigger Channel, Cable Select, and Trigger Ratio
settings.
3.
Select the required measurement device function to measure edge response.
4.
Ensure that the 9500B is in HEAD CAL, Edge, 150 ps Edge, LIN mode.
5.
Set 9500B's output ON, selecting TARGET 1 from Table 39.
6.
Select a measurement device range that gives an on-scale reading.
7.
Adjust the 9500B's output amplitude to give a reading equal to Target Amplitude
on the oscilloscope.
8.
Press ACCEPT CALIB.
9.
Select the next TARGET in Table 39 by pressing NEXT TARGET and return to
step 4; repeat until no TARGETs remain.
Table 39. Edge Function: 150 ps Edge: Linearity
Cal. Point
Voltage
Frequency
Edge
Target 1
1.0000 V
100 kHz
Rising
Target 2
2.0000 V
100 kHz
Rising
Target 3
2.5000 V
100 kHz
Rising
Target 4
1.0000 V
100 kHz
Falling
Target 5
2.0000 V
100 kHz
Falling
Target 6
2.5000 V
100 kHz
Falling
Calibration Procedure: 150 ps Edge: Gain
1.
Ensure that the 9500B is connected to the oscilloscope as shown in Figure 20,
and that both instruments are powered on and warmed up.
2.
Select the required 9500B Trigger Channel, Cable Select, and Trigger Ratio
settings.
3.
Select the required measurement device function to measure edge response.
4.
Ensure that the 9500B is in HEAD CAL, Edge, 150 ps Edge, GAIN mode.
5.
Set 9500B's output ON, selecting TARGET 1 from Table 40.
6.
Select a measurement device range that gives an on-scale reading.