High Performance Oscilloscope Calibrator
9510/9530/9550/9560 Head Calibration Procedures
119
Equipment Requirements
•
The UUT Active Head, connected to a verified Model 9500B Mainframe.
•
A high resolution Standards DMM with RMS AC Voltage accuracy of ±0.3 % or
better, from 10 mV to 3 V at 1 kHz.
Example: Model 1281 Digital Multimeter.
•
An adaptor to convert from BNC or PC3.5 to 4 mm leads.
Example: Model 4955 Calibration Adaptor.
•
Short, high-quality 4 mm leads for connection between the Calibration Adaptor
the DMM.
Calibration Setup
Connections: Ensure that the 9500B is connected to the DMM as shown in Figure 18,
or via a similar BNC or PC3.5 to 4 mm adaptor, and that both instruments
are powered on and warmed up.
erw238
Figure 18. LF Sine Voltage Calibration — Interconnections